PLOT Column Head Pressure Chart
Length |
Diameter |
Coating Thickness |
Fused Silica Part # |
MXT Part # |
Approx. Head Pressure (psi) |
Head Pressure (kpa) |
|
Q-BOND |
15m |
0.25mm |
8µm |
6.0 |
41 |
||
30m |
0.25mm |
8µm |
9.7 |
67 |
|||
15m |
0.32mm |
10µm |
3.4 |
23 |
|||
30m |
0.32mm |
10µm |
565926 |
5.0 |
34 |
||
15m |
0.53mm |
20µm |
1.3 |
9 |
|||
30m |
0.53mm |
20µm |
2.5 |
17 |
|||
50m |
0.53mm |
20µm |
565436 |
3.5 |
24 |
Length |
Diameter |
Coating Thickness |
Fused Silica Part # |
MXT Part # |
Approx. Head Pressure (psi) |
Head Pressure (kpa) |
|
QS-BOND |
15m |
0.25mm |
8µm |
7.0 |
48 |
||
30m |
0.25mm |
8µm |
14.0 |
97 |
|||
15m |
0.32mm |
10µm |
2.5 |
17 |
|||
30m |
0.32mm |
10µm |
5.3 |
37 |
|||
15m |
0.53mm |
20µm |
1.2 |
8 |
|||
30m |
0.53mm |
20µm |
2.2 |
15 |
Length |
Diameter |
Coating Thickness |
Fused Silica Part # |
MXT Part # |
Approx. Head Pressure (psi) |
Head Pressure (kpa) |
|
S-BOND |
15m |
0.25mm |
8µm |
6.0 |
41 |
||
30m |
0.25mm |
8µm |
9.3 |
64 |
|||
15m |
0.32mm |
10µm |
3.0 |
21 |
|||
30m |
0.32mm |
10µm |
6.0 |
41 |
|||
15m |
0.53mm |
20µm |
1.4 |
10 |
|||
30m |
0.53mm |
20µm |
2.6 |
18 |
Length |
Diameter |
Coating Thickness |
Fused Silica Part # |
MXT Part # |
Approx. Head Pressure (psi) |
Head Pressure (kpa) |
|
U-BOND |
15m |
0.25mm |
8µm |
6.0 |
41 |
||
30m |
0.25mm |
8µm |
9.5 |
66 |
|||
15m |
0.32mm |
10µm |
565658 |
3.6 |
25 |
||
30m |
0.32mm |
10µm |
6.3 |
43 |
|||
15m |
0.53mm |
20µm |
1.5 |
10 |
|||
30m |
0.53mm |
20µm |
2.5 |
17 |
Length |
Diameter |
Coating Thickness |
Fused Silica Part # |
MXT Part # |
Approx. Head Pressure (psi) |
Head Pressure (kpa) |
|
Msieve 5A |
15m |
0.25mm |
20µm |
10.0 |
69 |
||
15m |
0.32mm |
30µm |
565659 |
6.3 |
43 |
||
30m |
0.32mm |
30µm |
9.5 |
66 |
|||
15m |
0.53mm |
50µm |
2.0 |
14 |
|||
30m |
0.53mm |
50µm |
3.6 |
25 |
|||
50m |
0.53mm |
50µm |
564277 |
6.8 |
47 |
Length |
Diameter |
Coating Thickness |
Fused Silica Part # |
MXT Part # |
Approx. Head Pressure (psi) |
Head Pressure (kpa) |
|
Alumina BOND (Na2SO4) |
30m |
0.25mm |
4µm |
11.6 |
80 |
||
15m |
0.32mm |
5µm |
564952 |
6.6 |
46 |
||
30m |
0.32mm |
5µm |
565166 |
8.9 |
61 |
||
30m |
0.32mm |
3µm |
564896 |
9.0 |
62 |
||
50m |
0.32mm |
5µm |
15.3 |
105 |
|||
30m |
0.53mm |
10µm |
3.3 |
23 |
|||
50m |
0.53mm |
10µm |
5.6 |
39 |
Length |
Diameter |
Coating Thickness |
Fused Silica Part # |
MXT Part # |
Approx. Head Pressure (psi) |
Head Pressure (kpa) |
|
Alumina BOND (KCl) |
30m |
0.25mm |
4µm |
11.6 |
80 |
||
30m |
0.32mm |
5µm |
8.6 |
59 |
|||
50m |
0.32mm |
5µm |
14.6 |
101 |
|||
30m |
0.53mm |
10µm |
3.3 |
23 |
|||
50m |
0.53mm |
10µm |
5.4 |
37 |
Length |
Diameter |
Coating Thickness |
Fused Silica Part # |
MXT Part # |
Approx. Head Pressure (psi) |
Head Pressure (kpa) |
|
Alumina BOND (CFC) |
50m |
0.32mm |
5µm |
565682 |
11.2 |
77 |
|
30m |
0.53mm |
10µm |
2.6 |
18 |