Seven EPA Methods on One Column Pair using a µ-ECD

Seven EPA Methods on One Column Pair using a µ-ECD

Analyze Pesticides, PCBs, Herbicides and More on a Single Rtx®-CLPesticides Column Set

Download PDF

Get improved resolution, faster analyses, and less downtime by using Rtx®-CLPesticides and Rtx®-CLPesticides2 columns for multiple GC/ECD methods. Instead of changing columns between methods, you can save time by analyzing chlorinated pesticides, PCBs, herbicides, and other halogenated compounds on a single column set using an Agilent µ-ECD. Rtx®-CLPesticides and Rtx®-CLPesticides2 columns have a unique selectivity and outperform other column sets offered specifically for multiple GC/ECD methods.

EPA Method Column Pair Analysis Time (min.) Coelutions Restek Advantage
Comparison based on published competitor data. NDP = no data published
8081B
(Organochlorine pesticides)
Rtx-CLPesticides /
Rtx-CLPesticides2
7 / 7 0 / 0
DB-35ms / DB-XLB 15 / 16 0 / 0
ZB-MR1 / ZB-MR2 10 / 9 0 / 0
8081B (extended)(Organochlorine pesticides) Rtx-CLPesticides /
Rtx-CLPesticides2
24 / 23 1 / 2
  • Best balance of speed and selectivity.
  • All compounds are resolved between both columns.
  • See more info.
DB-35ms / DB-XLB 42 / 39 2 / 3
ZB-MR1 / ZB-MR2 NDP / 16 NDP / 3
8082A
(Polychlorinated biphenyls [PCBs], Aroclors)
Rtx-CLPesticides /
Rtx-CLPesticides2
7 / 7 0 / 0
  • Analyze PCBs 2x or 3x faster than on other ECD columns.
  • See more info.
DB-35ms / DB-XLB 14 / 16 0 / 0
ZB-MR1 / ZB-MR2 24 / 21 0 / 0
8151A
(Chlorinated herbicides)
Rtx-CLPesticides /
Rtx-CLPesticides2
13 / 13 1 / 0
DB-35ms / DB-XLB 16 / 17 0 / 0
ZB-MR1 / ZB-MR2 16 / 15 1 / 1
504.1
(EDB, DBCP, TCP)
Rtx-CLPesticides /
Rtx-CLPesticides2
6 / 6 0 / 0
  • Reliably separate analytes from trihalomethane interferences.
  • See more info.
DB-35ms / DB-XLB NDP NDP
ZB-MR1 / ZB-MR2 NDP NDP
505
(Organohalide pesticides)
Rtx-CLPesticides /
Rtx-CLPesticides2
18 / 18 1 / 1
DB-35ms / DB-XLB NDP NDP
ZB-MR1 / ZB-MR2 NDP NDP
508.1
(Chlorinated pesticides, herbicides, organohalides)
Rtx-CLPesticides /
Rtx-CLPesticides2
23 / 24 2 / 2
  • All compounds resolved between both columns.
  • Best overall balance of speed and resolution.
  • See more info.
DB-35ms / DB-XLB 22 / 24 2 / 4
ZB-MR1 / ZB-MR2 18 / NDP 2 / NDP
552.2
(Haloacetic acids, dalapon)
Rtx-CLPesticides /
Rtx-CLPesticides2
12 / 12 0 / 0
  • No coelutions—get accurate results for compounds that coelute on other columns.
  • See more info.
DB-35ms / DB-XLB 8 / 9 2 / 1
ZB-MR1 / ZB-MR2 NDP NDP

Download PDF

Get everything you need in one quick order—only Restek offers columns, standards, and sample prep supplies!