Excellent Responses in GC/MS Analysis of Semivolatiles
For Low Level GC/MS Analysis
- Inert, low-bleed column for reliable results from low-level GC/MS analyses.
- Save time — analyze acidic and basic compounds under the same conditions.
- Guaranteed reproducible performance, column to column.
The second column in our new Rxi® GC column line — the Rxi®-1ms column — will provide the same outstanding performance as the Rxi®-5ms column, with equally superior inertness, ultra-low bleed, and excellent batch to batch reproducibility.
Our first test for this 100% dimethylpolysiloxane phase column was an analysis of a complex mixture of semivolatile organic compounds. The extensive target list was comprised of many classes of compounds including chloroacetanilides, chlorotriazines, triazinones, uracils, polycyclic aromatic hydrocarbons, and phthalates. Figure 1 shows peak shape and selectivity are equally good for all of these diverse compounds, and all are eluted in an acceptable analysis time.1
Excellent Inertness
In addition to analyzing these compounds, we analyzed an acidic compound (2,4-dinitrophenol) and a basic compound (pyridine), each at 0.5ng on column, to assess column inertness. Column activity reveals itself through poor response and peak tailing for such active compounds, and these two compounds present both varying difficulties in a GC/MS analysis and differing modes of degradation. Figure 2 shows the excellent peak shapes and responses for these compounds on the 30m x 0.25mm ID, 0.25µm film column.
Phenols are notorious for breakdown and peak tailing, caused by interaction with the surface of an active inlet liner or an active column. Nitrophenols and pentachlorphenol, for example, very often exhibit poor peak shape and/or poor response. Figure 3 shows the 30m x 0.25mm ID, 0.25µm Rxi®-1ms column provides very good peak shapes for phenols. Peak responses are well above method requirements.
Ultra-Low Bleed
In addition to excellent inertness, Rxi®-1ms columns exhibit very low bleed. Figure 4 is focused on the end of the chromatogram for semivolatiles. At 330°C, bleed is much lower than the signals for 0.5ng of target analytes. This exceptional signal-to-noise differential for late eluting compounds assures better detection limits.
Based on these results, we highly recommend the new Rxi®-1ms column for low-level analyses that require a 100% dimethylpolysiloxane phase.
| Figure 1 Excellent selectivity and peak shapes for common drinking water semivolatiles at 10ng, using an | |||||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| |||||||||||||||||||||||||||||||||||||
| Figure 2 An Rxi®-1ms column has excellent selectivity for basic or acidic compounds, under the same conditions (0.5ng each; extracted ion chromatograms). | |||
|---|---|---|---|
| |||
| Figure 3 Symmetric peaks for acidic analytes at 5.0ng on an Rxi®-1ms column (extracted ion chromatogram). | ||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
![]()
| ||||||||||||||||||||||||
| Figure 4 Negligible bleed for an Rxi®-1ms column at 330°C (end of analysis for 5ng on-column standard). | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| |||||||||
The Drilled Uniliner® (Footnote 1)
To reduce the effects of surface activity in the injection port liner, and focus on the effects of the column on active analytes, we used a Drilled Uniliner® inlet liner for this work. This liner connects directly to the column, eliminating contact between the active compounds and active metal surfaces in the injector, and ensuring an inactive sample pathway for analyte transfer from the injection port to the column.

