High Throughput of Semivolatile Samples by GC/MS
Using a 20m x 0.18mm x 0.30µm Rxi®-5ms Column
- Increase productivity and sample throughput by 75%.
- Improved resolution of benzo(b)fluoranthene & benzo(k)fluoranthene.
- Highly inert, ultra-low bleed column is ideal for GC/MS analyses.
Analysis of semivolatile contaminants is complicated by variation in matrix type and complexity, as well as by the broad variety of chemicals tested. Extensive lists of up to 100 compounds are routinely analyzed, and potential target compounds can include aldehydes, ketones, anilines, phenols, ethers, pyridines, polynuclear aromatic hydrocarbons, and multiple types of pesticides. Coupled with the inherent complications of difficult analyses and problematic matrices, most laboratories also are challenged by trying to increase sample throughput.
Restek developed the 20m x 0.18mm x 0.30µm Rxi®-5ms capillary column to balance the demand for shorter analysis times with the need to maintain both column capacity and column efficiency. Analysis times for semivolatiles on typical 30m x 0.25mm x 0.5µm, 5% diphenyl/95% dimethyl polysiloxane columns generally range from 19 to 24 minutes. Using the short, thin film Rxi®-5ms capillary column, we have established conditions for eluting more than 90 semivolatiles, including 7 surrogates and 6 internal standards, in less than 12 minutes (Figure 1).
The new 20m x 0.18mm x 0.30µm Rxi®-5ms capillary column enables us to use an initial temperature of 55°C. By starting the analysis at a temperature 15°C higher than that used with typical 30m x 0.25mm x 0.5µm columns, we cut the oven cool-down time to 6 minutes, and reduced total analysis cycle time from 31 to 18 minutes. Cycle times were further reduced by eliminating the oven equilibrium hold time feature since the autosampler syringe rinse procedure provided an adequate interval between runs (saving another 0.5 minute/run).
Many methods, US EPA Method 8270 for example, require that specific quality assurance measures be met every 12 hours. Assuming all the required system performance checks and calibration verifications pass, there may only be a 10.5-hour window for sample analysis. Using a 20m x 0.18mm x 0.30µm Rxi®-5ms column under the conditions listed for Figure 1, a typical lab should be able to process 35 sample extracts, rather than an anticipated 20 extracts with a 30m x 0.25mm x 0.5µm column, in the same 10.5 hour analysis window. The additional 15 sample extracts per 12 hour shift represent a 75% increase in sample throughput (Table 1). We recommend you try a 20m x 0.18mm x 0.30µm Rxi®-5ms column, and see how much you can reduce your cost per analysis.
Table I Analyze 75% more samples with a 20m x 0.18mm x 0.30µm Rxi®-5ms capillary column. | ||||||
|---|---|---|---|---|---|---|
Restek 20m x 0.18mm x 0.30µm Rxi®-5ms column | 30m x 0.25mm x 0.5µm, 5% diphenyl/95% dimethyl polysiloxane column | Evaluation | ||||
Total run time (min.): | 18 | 31 | 13 minute savings/run | |||
Sample analysis (min.): | 12 | 22 | ||||
Oven cooling (min.): | 6 | 9 | ||||
Number of samples analyzed: | 35 | 20 | 75% increase | |||
(Data based on a 10.5 hour sample analysis period, following quality assurance and method performance samples, in a 12 hour shift.) | ||||||
| Figure 1 Analyze more than 90 semivolatiles in 12 minutes, using a 20m x 0.18mm x 0.30µm Rxi®-5ms column. | ||||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| ||||||||||||||||||||||||||||||||||||
