Fast, Accurate FAMEs Analyses of Biodiesel Fuel
Using a Stabilwax® Capillary GC Column
- Stable baselines, excellent peak symmetry, baseline resolution of all compounds.
- Analysis complete in less than 11 minutes.
- All RSD% values less than 1%.
A Stabilwax® fused silica GC column affords excellent peak symmetry, resolution, and reproducibility for determining the fatty acid methyl ester (FAME) and linolenic acid methyl ester content in B100 biodiesel fuel, using European standard method EN 14103. The chromatograms and quantified data shown here were generated from four different sources of biodiesel fuel, and meet or exceed the method criteria.
As biodiesel fuel continues to stimulate interest worldwide as an energy source, several gas chromatographic methods have been developed to determine the quality of B100 fuel. European standard method EN 14103 is used for determining the FAME and linolenic acid methyl ester content, European standard method EN 14105 and ASTM standard method D-6584-00e1 are used for determining free and total glycerin, and European standard method EN 14110 is used for determining residual methanol. Method EN 14103 permits the analyst to assure the B100 product is greater than 90% fatty acid methyl esters (m/m) and the linolenic acid content is between 1% and 15% (m/m). The analysis is appropriate for FAME compositions between C14:0 and C24:1.
In evaluating the suitability of the Stabilwax® column for quantifying FAMEs and linolenic acid methyl ester by method EN 14103, we prepared reference standards from each of the four B100 fuel sources soy, tallow, rapeseed, and yellow grease (Table 1) by weighing 250mg of the source material into a 10mL vial, then adding 5mL of a 10mg/mL solution of internal standard methyl heptadecanoate. (Avoid allowing the samples to stand longer than 12 hours, or quantification will be inaccurate.) We installed the 30m x 0.32mm ID x 0.25µm Stabilwax® column (cat.# 10624) in an Agilent 6890 instrument equipped with a split/splitless injector, a flame ionization detector, and ChemStation software. To obtain the fastest analysis, without sacrificing resolution, we selected hydrogen as the carrier gas, supplied from a Parker Balston hydrogen generator.
Figure 1 shows, for each source material, the analysis to FAME C24:1 is completed in less than 11 minutes. Particularly notable are the stability of the baselines, the excellent peak symmetry, and baseline resolution of all compounds of interest. Table 2 summarizes the RSD% values for the FAMEs measurements, all of which are less than 1%.
A 30m x 0.32mm ID x 0.25µm Stabilwax® column, used with hydrogen carrier gas, permits high speed analysis and ensures precise data acquisition for accurate quantification of C14:0-C24:1 FAMEs and linolenic acid methyl ester.
Table I Sources of FAMEs in B100 biodiesel fuel (% m/m). | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
Soy |
Tallow |
Rapeseed |
Yellow Grease |
|||||||
Myristic acid |
C14:0 |
0.21 |
1.7 |
0.11 |
0.68 |
|||||
Palmitic acid |
C16:0 |
11.24 |
25.5 |
4.1 |
16.35 |
|||||
Palmitoleic acid |
C16:1 |
0.2 |
3.27 |
0.27 |
1.23 |
|||||
Stearic acid |
C18:0 |
4.04 |
14.41 |
1.8 |
9.32 |
|||||
Oleic acid |
C18:1 |
21.93 |
40.34 |
58.57 |
47.8 |
|||||
Linoleic acid |
C18:2 |
53.84 |
12.02 |
22.2 |
20.01 |
|||||
Linolenic acid |
C18:3 |
7.29 |
0.99 |
13.26 |
2.93 |
|||||
Arachidic acid |
C20:0 |
0.36 |
0.4 |
0.79 |
0.46 |
|||||
Gadoleic acid |
C20:1 |
0.26 |
1.03 |
1.79 |
0.39 |
|||||
Behenic acid |
C22:0 |
0.45 |
0.57 |
0.44 |
||||||
Erucic acid |
C22:1 |
0.13 |
0.23 |
|||||||
Lignoceric acid |
C24:0 |
0.16 |
0.34 |
0.3 |
0.24 |
|||||
Nervonic acid |
C24:1 |
0.17 |
0.54 |
|||||||
Table 2 Relative standard deviations for FAMEs do not exceed 1% in analyses on a Stabilwax® column (n = 3). |
||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
Soy |
Tallow |
Rapeseed |
Yellow Grease |
|||||||
Myristic acid |
C14:0 |
0.33 |
0.42 |
0.24 |
0.36 |
|||||
Palmitic acid |
C16:0 |
0.04 |
0.06 |
0.02 |
0.04 |
|||||
Palmitoleic acid |
C16:1 |
0.23 |
0.17 |
0.19 |
0.09 |
|||||
Stearic acid |
C18:0 |
0.05 |
0.02 |
0.13 |
0.19 |
|||||
Oleic acid |
C18:1 |
0.02 |
0.3 |
0.2 |
0.25 |
|||||
Linoleic acid |
C18:2 |
0.25 |
0.41 |
0.11 |
0.22 |
|||||
Linolenic acid |
C18:3 |
0.13 |
0.16 |
0.07 |
0.14 |
|||||
Arachidic acid |
C20:0 |
0.3 |
0.37 |
0.23 |
0.31 |
|||||
Gadoleic acid |
C20:1 |
0.33 |
0.28 |
0.37 |
0.41 |
|||||
Behenic acid |
C22:0 |
0.28 |
0.29 |
0.17 |
||||||
Erucic acid |
C22:1 |
0.21 |
0.26 |
|||||||
Lignoceric acid |
C24:0 |
0.53 |
0.14 |
0.1 |
0.33 |
|||||
Nervonic acid |
C24:1 |
0.55 |
0.83 |
|||||||
Figure 1 Stable baselines, excellent peak symmetry, and rapid, baseline resolution of all compounds characterize FAMEs analyses on a Stabilwax® column. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Soy
IS Methyl heptadecanoate ![]() GC_PC00915
Tallow
IS Methyl heptadecanoate ![]() GC_PC00916
Rapeseed
IS Methyl heptadecanoate ![]() GC_PC00917
Yellow Grease
IS Methyl heptadecanoate ![]() GC_PC00918
|
Is your lab wasting money on bottled gas? |
|---|
If you use 2-3 cylinders of helium and/or hydrogen per week, as carrier gas and/or fuel gas, bottled gas is an expense in the range of $15,000 to $25,000 per year*, including overhead: expenses and time involved with ordering, transporting, installing, and periodically inspecting cylinders. You also contend with unquantifiable costs, such as floor space lost to an inventory of cylinders. Helium is a non-renewable resource extracted from natural gas and, because it is a petrochemical product, its cost will continue to rise, domestically and internationally. Chromatographers must look for cost effective, ultra-pure gas alternatives to supply their instruments and state-of-the-art analytical columns. Fortunately, we do have options. Relative to helium as the GC carrier gas, hydrogen from a gas generator reduces gas costs, cuts analysis time by 50%, and reduces temperatures needed for eluting analytes which increases column lifetime. Parker ChromGas® hydrogen generators are safe, convenient, reliable, and easy to use. For more information see the article "Parker PEM Hydrogen Generators". *Cost estimate for USA, in US $. |