US EPA 8330, a test method for determining trace amounts of 14 nitroaromatics, nitramines, and nitrate esters, was recently revised to include three new target analytes. The new method, EPA 8330B, includes nitroglycerine (NG), pentaerythritol tetranitrate, (PETN), and 3,5-dinitroaniline (3,5-DNA) and now covers 17 analytes that are commonly found in explosive and propellant residues. This method uses reversed phase HPLC and dual wavelength UV detection (210 & 254nm) in conjunction with a primary and a confirmation column.
We recently assessed the performance of our current column offerings relative to the elution order and retention times of the new analytes in the revised method. Separations on all columns were accomplished using a simple, isocratic water:methanol mobile phase (Table 1). The primary and confirmation columns that we currently recommend for the analysis of EPA 8330 are the Ultra C18 and Pinnacle™ II Biphenyl columns, respectively. Based on this work, this combination will continue to work well for the revised method, EPA 8330B, as shown by the chromatograms in Figures 1 and 2. Both columns provide excellent resolution of the EPA 8330B analytes and their differing selectivity provides a true confirmation analysis.
As an alternative to the Ultra C18/Pinnacle™ II Biphenyl combination, a Pinnacle™ II C18 column and a Pinnacle™ Cyano column work well together as a primary-confirmation column set. Another column of interest is the Allure® Biphenyl column. A high organic mobile phase was required on this column but the analysis was completed in approximately six minutes (Table 1).
Restek can provide the tools necessary to perform your explosives analysis including the recent changes to EPA Method 8330B. For this method, we recommend an Ultra C18 / Pinnacle™ II Biphenyl column pair, for dependable results for all target analytes.
Table 1 Retention times for EPA 8330B analytes on various Restek columns; new target analytes are shown in red, highlighted cells indicate coelution. |
||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
Primary Columns |
Confirmation Columns |
|||||||||||
Mobile |
H2O:MeOH |
50:50 |
44:56 |
45:55 |
50:50 |
20:80 |
||||||
Phase |
Flow |
1.5 mL/min. |
1.0 mL/min. |
1.2 mL/min. |
1.5 mL/min. |
1.5 mL/min. |
||||||
Analytes |
Pinnacle™ II C18 |
Ultra C18 |
Pinnacle™ II Biphenyl |
Pinnacle™ II Cyano |
Allure® Biphenyl |
|||||||
1 |
HMX |
2.29 |
3.38 |
2.76 |
18.65 |
1.61 |
||||||
2 |
RDX |
3.63 |
5.41 |
3.22 |
9.38 |
1.75 |
||||||
3 |
1,3,5-TNB |
4.89 |
7.39 |
14.54 |
4.78 |
5.69 |
||||||
4 |
1,3-DNB |
5.94 |
8.82 |
9.26 |
4.59 |
3.92 |
||||||
5 |
3,5-DNA |
6.63 |
9.71 |
5.73 |
6.34 |
2.30 |
||||||
6 |
tetryl |
6.97 |
9.71 |
16.12 |
11.47 |
4.42 |
||||||
7 |
NB |
6.97 |
9.88 |
6.31 |
3.80 |
2.79 |
||||||
8 |
2,4,6-TNT |
8.23 |
11.69 |
20.17 |
5.94 |
6.22 |
||||||
9 |
NG |
8.23 |
11.69 |
4.94 |
8.52 |
1.98 |
||||||
10 |
2-A-4,6-DNT |
8.94 |
12.05 |
7.43 |
7.24 |
2.50 |
||||||
11 |
4-A-2,6-DNT |
8.94 |
12.61 |
7.02 |
6.34 |
2.41 |
||||||
12 |
2,6-DNT |
9.73 |
13.27 |
12.36 |
5.10 |
4.09 |
||||||
13 |
2,4-DNT |
9.73 |
13.64 |
15.46 |
5.58 |
5.14 |
||||||
14 |
2-NT |
11.92 |
15.92 |
9.73 |
4.38 |
3.40 |
||||||
15 |
4-NT |
12.76 |
17.05 |
11.07 |
4.38 |
3.72 |
||||||
16 |
3-NT |
13.74 |
18.32 |
11.07 |
4.38 |
3.73 |
||||||
17 |
PETN |
16.13 |
21.08 |
10.43 |
17.24 |
2.67 |
||||||
|
Figure 1 Excellent resolution of EPA 8330B target analytes on the Ultra C18 column. |
||||||||||||||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|