Ordering Notes
Technical Resources

Selected Product

Cat.#:

List Price:$0.00



Select a Product
catalog #lengthiddfunits
19773
15m
0.25mm
20μm
ea.
19720
15m
0.32mm
30μm
ea.
19722
30m
0.32mm
30μm
ea.
19721
15m
0.53mm
50μm
ea.
19723
30m
0.53mm
50μm
ea.

GC Column Installation Checklist

Installation Checklist for GC Columns—Restek Innovations and Technical Services specialists have found this checklist to be a reliable sequence for avoiding problems when installing a capillary GC column.

Molecular sieve materials are very hydrophilic

Because molecular sieve materials are very hydrophilic, they will adsorb water from the sample or carrier gas. Water contamination can have a detrimental effect on peak symmetry and can reduce the resolution of all compounds. If water contamination occurs, reactivate your Rt®-Msieve 5A PLOT column by conditioning at 300°C with dry carrier gas flow for 3 hours.