Ordering Notes
Technical Resources

Selected Product

Cat.#:

List Price:$0.00



Select a Product
catalog #lengthiddfunits
19773
15 m
0.25 mm
20 μm
ea.
19720
15 m
0.32 mm
30 μm
ea.
19722
30 m
0.32 mm
30 μm
ea.
19721
15 m
0.53 mm
50 μm
ea.
19723
30 m
0.53 mm
50 μm
ea.

GC Column Installation Checklist

Installation Checklist for GC Columns—Restek Innovations and Technical Services specialists have found this checklist to be a reliable sequence for avoiding problems when installing a capillary GC column.

Molecular sieve materials are very hydrophilic

Because molecular sieve materials are very hydrophilic, they will adsorb water from the sample or carrier gas. Water contamination can have a detrimental effect on peak symmetry and can reduce the resolution of all compounds. If water contamination occurs, reactivate your Rt®-Msieve 5A PLOT column by conditioning at 300°C with dry carrier gas flow for 3 hours.