Welcome, Guest -  Log In |  Register  |  Shopping Cart

Polynuclear Aromatic Hydrocarbons US EPA Method 610

  1. naphthalene
  2. acenaphthylene
  3. acenaphthene
  4. fluorene
  5. phenanthrene
  6. anthracene
  7. fluoranthene
  8. pyrene
  1. benzo(a)anthracene
  2. chrysene
  3. benzo(b)fluoranthene
  4. benzo(k)fluoranthene
  5. benzo(a)pyrene
  6. indeno(1,2,3-cd)pyrene
  7. dibenzo(a,h)anthracene
  8. benzo(ghi)perylene

Column:
Rxi™-5ms, 30m, 0.25mm ID, 0.25µm (cat.# 13423)
Sample:
SV Calibration Mix #5 / 610 PAH Mix (cat.# 31011)
Inj.:
1.0µL, 10ppm each analyte (10ng on column), splitless (hold 0.1 min.) 4mm Drilled Uniliner® inlet liner (hole at bottom) (cat.# 20771)
Instrument:
Agilent 6890
Inj. temp.:
275°C
Carrier gas:
helium, constant flow
Flow rate:
1.2mL/min.
Oven temp.:
75°C (hold 0.5 min.) to 245°C @ 25°C/min., to 330°C @ 4°C/min. (hold 1 min.)
Det.:
Agilent 5973 GC/MS
Transfer line temp.:
280°C
Scan range:
35-550 amu
Solvent delay:
2 min.
Tune:
DFTPP
Ionization:
EI

GC_EV00824