US EPA TO-15 Compounds on Rtx-1

US EPA TO-15 Compounds on Rtx®-1

Peaks
1.Propylene
2.Freon®-12 (dichlorodifluoromethane)
3.Chloromethane
4.Freon®-114 (dichlorotetrafluoroethane)
5.Vinyl chloride
6.1,3-Butadiene
7.Bromomethane
8.Chloroethane
9.Carbon disulfide
10.Acetone
11.Freon®-11 (trichlorofluoromethane)
12.Isopropyl Alcohol
13.1,1-Dichloroethene
14.Methylene Chloride
15.Freon®-113 (1,1,2-trichloro-1,2,2-trifluoroethane)
16.trans-1,2-Dichloroethene
17.1,1-Dichloroethane
18.Methyl tert-butyl ether
19.Vinyl acetate
20.Methyl ethyl ketone
21.cis-1,2-Dichloroethene
22.Hexane
23.Chloroform
24.Ethyl Acetate
25.Tetrahydrofuran
26.1,2-Dichloroethane
27.1,1,1-Trichloroethane
28.Benzene
29.Carbon Tetrachloride
30.Cyclohexane
Peaks
31.1,2-Dichloropropane
32.Trichloroethylene
33.Bromodichloromethane
34.1,4-Dioxane
35.Heptane
36.cis-1,3-Dichloropropene
37.Methyl Isobutyl Ketone
38.trans-1,3-Dichloropropene
39.1,1,2-Trichloroethane
40.Toluene
41.Methyl butyl ketone
42.Dibromochloromethane
43.1,2-Dibromoethane
44.Tetrachloroethylene
45.Chlorobenzene
46.Ethylbenzene
47.p-Xylene
48.m-Xylene
49.Bromoform
50.Styrene
51.o-Xylene
52.1,1,2,2-Tetrachloroethane
53.4-Ethyltoluene
54.1,3,5-Trimethylbenzene
55.1,2,4-Trimethylbenzene
56.1,3-Dichlorobenzene
57.Benzyl chloride
58.1,4-Dichlorobenzene
59.1,2-Dichlorobenzene
60.1,2,4-Trichlorobenzene
61.Hexachloro-1,3-Butadiene
image
GC_AR00748
ColumnRtx®-1, 60 m, 0.32 mm ID, 1.00 µm (cat.# 10157)
SampleTO-15 65 Component Mix (cat.# 34436) humidified to 33% RH in a 6L SilcoCan® (cat.# 24182)
InjectionDirect
Oven
Oven Temp.:30 °C (hold 4 min) to 175 °C at 8 °C/min to 220 °C at 20 °C/min (hold 2 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:150 °C
Analyzer Type:Quadrupole
Ionization Mode:EI
Scan Range:35-280 amu
PreconcentratorNutech 3550A Preconcentrator
  Trap 1 Settings
Cooling temp:-160 °C
Desorb temp:150 °C
  Cryofocuser
Cooling temp:-185 °C
Desorb temp:150 °C
  Standard
Size:300 mL
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

TO-15

,

toxic organic

,

air

,

volatiles

,

xylene

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.