Mass APH on Rxi®-1ms

Peaks
1.1,3-Butadiene
2.iso-Pentane
3.Methyl tert-butyl ether
4.Bromochloromethane
5.n-Hexane
6.Benzene
7.Cyclohexane
8.1,4-Difluorobenzene
9.2,3-Dimethylpentane
10.n-Heptane
11.Toluene
12.n-Octane
13.Chlorobenzene-d5
14.Ethylbenzene
Peaks
15.2,3-Dimethylheptane
16.m-Xylene/p-Xylene
17.o-Xylene
18.n-Nonane
19.4-Bromofluorobenzene
20.Isopropylbenzene
21.1-Ethyl-3-methylbenzene
22.1,3,5-Trimethylbenzene
23.n-Decane
24.1,2,3-Trimethylbenzene/p-Isopropyltoluene
25.Butylcyclohexane
26.n-Undecane
27.Naphthalene
28.n-Dodecane
image
GC_AR1145
ColumnRxi®-1ms, 60 m, 0.32 mm ID, 1.00 µm (cat.# 13357)
Injectionsplitless
Oven
Oven Temp:35 °C (hold 5 min.) to 220 °C at 7 °C/min. (hold 2 min.)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min.
DetectorMS
Mode:Scan
Transfer Line Temp.:240 °C
Analyzer Type:Quadrupole
Source Temp.:230 °C
Quad Temp.:150 °C
Solvent Delay Time:4.0 min.
Ionization Mode:EI
Scan Range:35-250 amu
PreconcentratorEntech 7100
InstrumentHP6890 GC & 5973 MSD
AcknowledgementCourtesy of Mass DEP, Northeast Region