Derivatized Amphetamines (500 ng/mL) on Rxi®-5Sil MS

Peaks
1.Amphetamine
2.Methamphetamine
Peaks
3.MDA
4.MDMA
5.MDEA
c = contaminant
image
GC_CF1144
ColumnRxi®-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
Sample
Diluent:Butyl chloride
Conc.:500 ng/mL HFAA derivatives
Injection
Inj. Vol.:1 µL splitless (hold 1 min.)
Liner:3.5mm Gooseneck Splitless w/Wool (cat.# 22286-200.1)
Inj. Temp.:250 °C
Purge Flow:28 mL/min.
Oven
Oven Temp:75 °C to 300 °C at 15 °C/min.
Carrier GasHe, constant linear velocity
Linear Velocity:45 cm/sec., 13.5 psi, 93.1 kPa @ 75 °C
DetectorMS
Mode:Scan
Transfer Line Temp.:250 °C
Analyzer Type:Quadrupole
Source Temp.:200 °C
Electron Energy:70 eV
Solvent Delay Time:4 min.
Tune Type:PFTBA
Ionization Mode:EI
Scan Range:40-300 amu
Scan Rate:5 scans/sec.
InstrumentShimadzu 2010 GC & QP2010+ MS

RELATED CHROMATOGRAM LIBRARY SEARCHES

Amphetamines

,

GCMS

,

Rxi-5

,

Rxi-5MS

,

Rxi-5SilMS