share:
delim
delim
Solvents on Rtx-Volatiles

Solvents on Rtx-Volatiles

Peaks
1.Water
2.Methanol
3.Ethanol
4.Isopropanol
5.Ethyl acetate
6.Isopropyl acetate
7.4-Methyl-2-pentanone
image
GC_CH00352
ColumnRtx-Volatiles, 60 m, 0.53 mm ID, 2.00 µm (cat.# 10905)
Sampleneat solvents
Injection
Inj. Vol.:1.0 µL split (split ratio 40:1)
Inj. Temp.:250 °C
Oven
Oven Temp.:40 °C (hold 2 min) to 200 °C at 5 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:2.6 mL/min
Linear Velocity:20 cm/sec
DetectorTCD @ 250 °C
NotesTCD sensitivity: 4 x 10-11 AFS
TCD current: 100mA
Download PDF