share:
delim
delim
Dioxins (HxCDFs) in Fly Ash on Rxi-5Sil MS

Dioxins (HxCDFs) in Fly Ash on Rxi-5Sil MS

image
GC_EV1199
ColumnRxi-5Sil MS, 60 m, 0.18 mm ID, 0.10 µm (cat.# 43607)
SampleFly ash extract
Diluent:Nonane
Injection
Inj. Vol.:1 µL splitless
Liner:2 mm splitless liner (cat.# 20712)
Oven
Oven Temp.:120 °C (hold 1 min) to 160 °C at 10 °C/min to 300 °C at 2.5 °C/min
Carrier GasHe, constant flow
Flow Rate:1 mL/min
DetectorWaters AutoSpec Ultima Mass Spectrometer
Source Temp.:280 °C
Ionization Mode:EI
NotesElectron Ionization at 40eV
Red indicates toxicity.
AcknowledgementChromatogram courtesy of Karen MacPherson, Li Shen, Terry Kolic, and Eric Reiner at the Ontario Ministry of the Environment
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Hexachlorodibenzofuran

,

dioxin

,

para dioxin

,

p dioxin

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.