EPA 522 Analyte Mix on Rxi®-624Sil MS (30 m x 0.25 mm x 1.4 μm)
| Peaks | RT (min) |
| 1. | Tetrahydrofuran-D8 | 5.27 |
| 2. | 1,4-Dioxane-d8 | 6.19 |
| 3. | 1,4-Dioxane | 6.23 |
GC_EV1251
| Rxi®-624Sil MS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 13868) |
| using Rxi® guard column 5 m, 0.25 mm ID (cat.# 10029) |
| with Universal Angled Press-Tight® connectors (cat.# 20446-261) |
| Tetrahydrofuran-d8 (cat.# 30112) |
| 1,4-Dioxane-d8 (cat.# 30614) |
| 1,4-Dioxane (cat.# 30287) |
| Dichloromethane |
| 50 ng/mL |
|
| 10 µL splitless (hold 1 min) |
| Sky™ 4 mm single taper/gooseneck w/wool (cat.# 23303.5) |
| 120 °C |
| 80 mL/min |
|
| 35 °C (hold 1 min) to 120 °C at 12 °C/min (hold 1 min) |
| He, constant flow |
| 1.4 mL/min |
| 30.556 cm/sec @ 35 °C |
| MS |
| SIM |
| Group | Start Time (min) | Ion(s) | Dwell (ms) |
|---|
| 1 | 5.0 | 46,78,80 m/z | 50 | | 2 | 5.85 | 96,88,64,62,58 m/z | 40 |
|
| 280 °C |
| Quadrupole |
| 230 °C |
| 150 °C |
| 5.0 min |
| BFB |
| EI |
| Agilent 7890A GC & 5975C MSD |
| Large volume splitless injection for EPA Method 522 without a PTV. |
| M. Misselwitz, J. Cochran |

RELATED CHROMATOGRAM LIBRARY SEARCHES