Electronic Cigarette Liquid on Rtx-VMS (Hydrogen)

Electronic Cigarette Liquid on Rtx®-VMS (Hydrogen)

PeakstR (min)
1.Methanol0.875
2.Ethanol0.926
3.Methylene chloride0.987
4.Propylene glycol1.619
5.Glycerin2.805
6.Nicotine3.927
image
GC_FF1258
ColumnRtx®-VMS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 19915)
Sample
Diluent:Methylene chloride
Conc.: Electronic cigarette liquid diluted 100:1
Injection
Inj. Vol.:1.0 µL split (split ratio 200:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.5)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 240 °C at 35 °C/min
Carrier GasH2, constant flow
Flow Rate:2.5 mL/min
Linear Velocity:67.2 cm/sec @ 100 °C
DetectorFID @ 250 °C
Make-up Gas Flow Rate:50 mL/min
Make-up Gas Type:H2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
InstrumentAgilent 7890A GC