share:
delim
delim
Cannabinoids in Cannabis Sample 10 on Rxi-35Sil MS

Cannabinoids in Cannabis Sample 10 on Rxi®-35Sil MS

PeakstR (sec)Conc. (%)
1.Cannabichromene82.0-
2.Cannabidiol84.6-
3.delta-8-Tetrahydrocannabinol93.5-
4.delta-9-Tetrahydrocannabinol97.18.3%
5.Cannabigerol99.4-
6.Cannabinol104.9-
image
GC_FS0528
ColumnRxi®-35Sil MS, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13820)
SampleCannabis sample 10
Conc.:
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Premium 4 mm Precision® liner w/wool (cat.# 23305.5)
Inj. Temp.:250 °C
Split Vent Flow Rate:25 mL/min
Oven
Oven Temp.:225 °C (hold 0.1 min) to 330 °C at 35 °C/min (hold 0.9 min)
Carrier GasH2, constant flow
Flow Rate:2.5 mL/min
DetectorFID @ 350 °C
Constant Column + Constant Make-up:50 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Medical cannabis

,

Cannabidiol

,

Cannabichromene

,

Cannabigerol

,

delta-9-THC

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.