share:
delim
delim
Gasoline Containing Ethanol by ASTM D3606-06 (Modified) on D3606 Application Column Set

Gasoline Containing Ethanol by ASTM D3606-06 (Modified) on D3606 Application Column Set

Peaks
1.Ethanol
2.Benzene
3.2-Butanol (IS)
4.Toluene
image
GC_PC01079
ColumnD3606 Application Column (2 column set) Column 1: 6' (1.8m), 1/8" OD, 2.0mm ID, nonpolar Rtx®-1 Column 2: 16' (4.9m), 1/8" OD, 2.0mm ID, proprietary packing material (cat.# 83606-800)
SampleGasoline with internal standard
Diluent:
Injectionsample valve
Sample Loop Vol.:1.5 µL
Oven
Oven Temp.:135 °C (hold 16 min)
Carrier GasHe, constant flow
Flow Rate:20.0 mL/min
DetectorTCD @ 200 °C
Notes3 min. backflush
AcknowledgementChromatogram courtesy of Boguslaw Dudek, Conoco Phillips, Linden, NJ
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

packed column

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.