Simulated Distillation on MXT-1HT SimDist by ASTM D7213 Using Nitrogen Carrier Gas

Simulated Distillation on MXT®-1HT SimDist by ASTM D7213 Using Nitrogen Carrier Gas

image
GC_PC1286
ColumnMXT®-1HT SimDist, 5 m, 0.53 mm ID, 0.88 µm (cat.# 70131)
SampleRetention time standard for D7213 (C5-C72), Separation Systems, Inc.
Diluent:Carbon disulfide
Conc.:1 wt%
Injection
Inj. Vol.:1 µL PTV on column
Liner:Premium 1.7 mm ID on-column inlet liner (cat.# 23430.1)
Inlet Temp. Program:90 °C to 380 °C at 20 °C/min
Oven
Oven Temp.:40 °C to 380 °C at 10 °C/min (hold 10 min)
Carrier GasN2, constant flow
Flow Rate:11.9 mL/min
DetectorFID @ 390 °C
Make-up Gas Flow Rate:18 mL/min
Make-up Gas Type:N2
Hydrogen flow:43 mL/min
Air flow:400 mL/min
InstrumentAgilent 7890B GC