Fatty Acids (C1-C7) on DiatoSorb-WHP Packed Column

Fatty Acids (C1-C7) on DiatoSorb-WHP Packed Column

Peaks
1.Acetic acid
2.Formic acid
3.Propionic acid
4.Isobutyric acid
5.n-Butyric acid
6.Isovaleric acid
7.n-Valeric acid
8.Isocaproic acid
9.n-Caproic acid
10.Heptanoic acid
image
GC_PC1289
Column10% FFAP, 1% H3PO4, 100/120 mesh on DiatoSorb-WHP, SilcoSmooth tubing, 2 m, 1/8 in OD, 2 mm ID
SampleFatty acids C1-C7
Diluent:Ether
Conc.:1,000 µg/mL
Injection
Inj. Vol.:8 µL on-column
Inj. Temp.:200 °C
Oven
Oven Temp.:125 °C (hold 2 min) to 175 °C at 10 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:30 mL/min
DetectorTCD @ 230 °C
InstrumentAgilent 7890B GC