share:
delim
delim
Simulated Distillation with Helium Carrier Gas on MXT-1HT SimDist by ASTM D2887-B (Accelerated)

Simulated Distillation with Helium Carrier Gas on MXT®-1HT SimDist by ASTM D2887-B (Accelerated)

image
GC_PC1303
ColumnMXT®-1HT SimDist, 10 m, 0.53 mm ID, 0.88 µm (cat.# 70134)
SampleASTM D2887-12 calibration standard (cat.# 31674)
Injection
Inj. Vol.:0.02 µL cool on-column
Temp. Program:100 °C to 360 °C at 35 °C/min
Oven
Oven Temp.:60 °C to 360 °C at 35 °C/min
Carrier GasHe, constant flow
Flow Rate:26 mL/min
DetectorFID @ 360 °C
Make-up Gas Flow Rate:20 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:360 mL/min
InstrumentAgilent 7890B GC
NotesAccelerated analysis based on ASTM Method D2887 (Procedure B)
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.