Simulated Distillation with Hydrogen Carrier Gas on MXT®-1HT SimDist by ASTM D2887-B (Accelerated)

GC_PC1304
Column | MXT®-1HT SimDist, 10 m, 0.53 mm ID, 0.88 µm (cat.# 70134) |
Sample | ASTM D2887-12 calibration standard (cat.# 31674) |
Injection | |
Inj. Vol.: | 0.02 µL cool on-column |
Temp. Program: | 80 °C to 360 °C at 35 °C/min |
Oven | |
Oven Temp.: | 60 °C to 360 °C at 35 °C/min |
Carrier Gas | H2, constant flow |
Flow Rate: | 22.35 mL/min |
Detector | FID @ 360 °C |
Make-up Gas Flow Rate: | 30 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 30 mL/min |
Air flow: | 400 mL/min |
Instrument | Agilent 7890B GC |
Notes | Accelerated analysis based on ASTM Method D2887 (Procedure B) |