share:
delim
delim
High-Temperature Simulated Distillation on MXT-1HT SimDist by ASTM D7169

High-Temperature Simulated Distillation on MXT®-1HT SimDist by ASTM D7169

image
GC_PC1308
ColumnMXT®-1HT SimDist, 5 m, 0.53 mm ID, 0.10 µm (cat.# 70112)
SampleRetention time standard for D7169 (C5-C100)
Injection
Inj. Vol.:0.1 µL cool on-column
Temp. Program:50 °C to 425 °C at 15 °C/min
Oven
Oven Temp.:-20 °C to 425 °C at 15 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:20 mL/min
DetectorFID @ 435 °C
Make-up Gas Flow Rate:15 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
InstrumentAgilent 7890B GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

High Temperature Gas Chromatography

,

Vacuum Residues

,

SimDis

,

Crude oil

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.