Permanent Gases with C2 and C3 Hydrocarbons on ShinCarbon ST

Permanent Gases with C2 and C3 Hydrocarbons on ShinCarbon ST

PeakstR (min)
1.Oxygen1.320
2.Nitrogen1.440
3.Carbon monoxide1.896
4.Methane3.839
5.Carbon dioxide6.265
6.Ethine (acetylene)8.587
7.Ethene (ethylene)9.271
8.Ethane10.092
9.Propadiene (allene)14.003
10.Propene (propylene)14.440
11.Propane14.889
*Elution of hydrogen
image
GC_PC1329
Column100/120 mesh on ShinCarbon ST, SilcoSmooth tubing, 2 m, 1.00 mm ID (cat.# 19808)
SamplePermanent gas standard with C2 and C3 hydrocarbons
Diluent:Helium
Conc.:1-5 mol% varied
Injection
Inj. Vol.:500 µL split (split ratio 5:1)
Liner:4.0 mm ID straight inlet liner (cat.# 20773)
Inj. Temp.:200 °C
Oven
Oven Temp.:40 °C (hold 2 min) to 280 °C at 20 °C/min (hold 3 min)
Carrier GasHe, constant flow
Flow Rate:15 mL/min
DetectorTCD @ 290 °C
InstrumentAgilent 7890B GC
NotesOxygen and nitrogen are baseline resolved.