Solvents on Rtx®-BAC1 & Rtx®-BAC2 (headspace analysis)
| Peaks |
| 1. | Ethyl chloride |
| 2. | Diethyl ether |
| 3. | Methylene chloride |
| 4. | Acetonitrile |
| 5. | Methyl tert-butyl ether |
| 6. | Chloroform |
| 7. | Methyl ethyl ketone |
| 8. | 1,1,1-Trichloroethane |
| Peaks |
| 9. | Benzene |
| 10. | Trichloroethylene |
| 11. | Toluene |
| 12. | 1,1,2-Trichloroethane |
| 13. | Tetrachloroethene |
| 14. | p-Xylene |
| 15. | m-Xylene |
| 16. | o-Xylene |
| 17. | Tetrachloroethane |
GC_PH00509_510
| Rtx®-BAC1 30 m, 0.53 mm ID, 3.00 µm (cat.# 18001) |
| and Rtx®-BAC2 30 m, 0.53 mm ID, 2.00 µm (cat.# 18000) |
|
|
| 1,000 µL headspace-syringe split |
| 240 °C |
|
| 40 °C (hold 5 min) to 240 °C at 5 °C/min |
| He, constant flow |
| 65 cm/sec |
| FID @ 240 °C |

RELATED CHROMATOGRAM LIBRARY SEARCHES