share:
delim
delim
Immunosuppressive Drugs in Blood on Raptor Biphenyl by LC-MS/MS
This chromatogram may contain analytes that are synonyms of your search terms.

Immunosuppressive Drugs in Blood on Raptor Biphenyl by LC-MS/MS

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion
1.Ascomycin1.4510809.53756.50
2.Tacrolimus1.4710821.61768.51
3.Sirolimus1.6410931.63864.57
4.Everolimus1.6910975.68908.62
5.Cyclosporin A1.83101219.831202.87
6.Cyclosporin D1.891001233.911216.88
image
LC_CF0686
ColumnRaptor Biphenyl (cat.# 9309A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Temp.:70 °C
Inj. Vol.:5 µL
Mobile Phase
A:0.05% Formic acid, 5 mM ammonium formate in water
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.54060
2.000.50100
2.010.54060
3.000.54060
Max Pressure:305 bar
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
NotesSample Preparation Procedure:
Human whole blood was fortified with 4 immunosuppressive drugs at 10 ng/mL. For quantitation, cyclosporin D was used as the internal standard for cyclosporin A and ascomycin was used as the internal standard for tacrolimus, sirolimus, and everolimus. The blood sample (100 μL) was mixed with 200 μL of precipitation solution (1:4 v/v 0.2 M ZnSO4:methanol) containing 50 ng/mL of cyclosporin D and 5 ng/mL of ascomycin. The mixture was vortexed for 20 seconds at 3,000 rpm and then centrifuged for 10 minutes at 4,300 rpm. The supernatant was directly injected (5 μL) for analysis.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

cyclosporin

,

tacrolimus

,

sirolimus

,

everolimus

,

immunosuppressive

,

immunosuppresant

,

raptor biphenyl

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.