16 Cannabinoids on Raptor ARC-18 1.8 μm by LC-UV

16 Cannabinoids on Raptor ARC-18 1.8 μm by LC-UV

PeakstR (min)
1.Cannabidivarinic acid (CBDVA)0.787
2.Cannabidivarin (CBDV)0.867
3.Cannabidiolic acid (CBDA)1.090
4.Cannabigerolic acid (CBGA)1.157
5.Cannabigerol (CBG)1.216
6.Cannabidiol (CBD)1.279
7.Tetrahydrocannabivarin (THCV)1.396
8.Tetrahydrocannabivarinic acid (THCVA)1.774
PeakstR (min)
9.Cannabinol (CBN)1.894
10.Cannabinolic acid (CBNA)2.257
11.Δ9-Tetrahydrocannabinol (Δ9-THC)2.380
12.Δ8-Tetrahydrocannabinol (Δ8-THC)2.456
13.Cannabicyclol (CBL)2.826
14.Cannabichromene (CBC)2.966
15.Tetrahydrocannabinolic acid A (THCA-A)3.150
16.Cannabichromenic acid (CBCA)3.516
image
LC_GN0579
ColumnRaptor ARC-18 (cat.# 931421E)
Dimensions:100 mm x 3.0 mm ID
Particle Size:1.8 µm
Pore Size:90 Å
Guard Column:UltraShield UHPLC precolumn filter, 0.2 μm frit (cat.# 25809)
Temp.:30 °C
SampleTetrahydrocannabivarin (cat.# 34100)
Cannabidiolic acid (CBDA) (cat.# 34099)
Cannabichromene (CBC) (cat.# 34092)
Cannabigerol (CBG) (cat.# 34091)
delta-9-Tetrahydrocannabinolic acid A (THCA-A) (cat.# 34093)
delta-8-Tetrahydrocannabinol (Δ8-THC) (cat.# 34090)
delta-9-Tetrahydrocannabinol (Δ9-THC) (cat.# 34067)
Cannabinol (CBN) (cat.# 34010)
Cannabidiol (CBD) (cat.# 34011)
Compounds not present in these mixes were obtained separately.
Diluent:25:75 Water:methanol
Conc.:50 µg/mL
Inj. Vol.:1 µL
Mobile Phase
A:Water, 5 mM ammonium formate, 0.1% formic acid
B:Acetonitrile, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.001.02575
4.001.02575
DetectorUV/Vis @ 228 nm
InstrumentUHPLC