Welcome, Guest -  Log In |  Register  |  Shopping Cart

Restek Capillary Column Installation Guide

Section I: Brief Installation Procedure

The following procedure is a quick, easy, trouble-free sequence for installing a capillary column. For more in-depth installation information, refer to Section II.

  1. Cool all heated zones.
  2. Replace spent oxygen and moisture traps.
  3. Clean and/or deactivate injector and detector sleeves if necessary.
  4. Replace critical injector and detector seals.
  5. Replace septum.
  6. Set make-up and detector gas flow rates.
  7. Carefully inspect the column for damage or breakage.
  8. Cut 10 centimeters from each end of the column.
    • use a sapphire scribe (cat.# 20115) or ceramic scoring wafer (cat.# 20116) to cut fused silica capillary columns.
    • use the serrated edge of a ceramic scoring wafer (cat.# 20116) or the edge of a sharp file to cut metal capillary columns.
  9. Install a nut and ferrule on each end of the column.
  10. Cut an additional 10 centimeters from each end of the column to remove ferrule fragments.
  11. Mount the capillary column in the oven using a bracket that protects the column from becoming scratched or abraded.
  12. Insert column the appropriate distance into the inlet as indicated in the instrument manual.
  13. Set the approximate column flow rate by adjusting the head pressure to the value listed on the test chromatogram included with the column.
  14. Set split vent, septa purge, and any other applicable inlet gases according to the instrument specifications.
  15. Confirm the flow by immersing the column outlet in a vial of solvent (acetone or isopropyl alcohol).
  16. Insert column the appropriate distance into the detector as indicated in the instrument manual.
  17. Check for inlet and outlet leaks using a thermal conductivity leak detector (cat.#'s 21607, 20130, or 21609). Do not use soaps or liquid-based leak detectors or the column may be damaged.
  18. Set injector and detector temperatures. Turn the detector on when the temperatures have equilibrated. Caution - do not exceed the phase's maximum operating temperature!
  19. To set the proper dead time (linear velocity), inject methane (cat.# 20197) or a non-retained substance compatible with the detector being used.
  20. Verify system integrity by checking the dead volume peak. It should not tail.
  21. Condition the column at its maximum operating temperature to stabilize the baseline. (See the test chromatogram included with the column for the maximum temperature.)
  22. Set oven to appropriate temperature and inject methane (cat.# 20197) or an appropriate unretained substance, again to set the proper linear velocity.
  23. Inject a duplicate of the original test mixture or your specific test mixture to confirm proper installation, system, and column performance.
  24. Calibrate the instrument and inject samples.
Restek Capillary Column Installation Guide
< prev next >