What gases will my Restek Electronic Leak Detector detect?

The Restek Electronic Leak Detector will detect minute leaks of any commonly used carrier, fuel, or make-up gas with a thermal conductivity different from air. The reference gas inlet draws in ambient air for comparison to air drawn into the sample probe. A thermal conductivity difference between the two indicates a leak, and the leak is revealed by both an LED display and an audible tone.


Gas
Min. Detectable
Leak Rate (atmcc/sec.)
Indicating LED
Light Color
Helium
1.0 x 10-5
Red
Nitrogen
1.4 x 10-3
Yellow
Argon
2.2 x 10-3
Yellow
Carbon dioxide
1.1 x 10-3
Yellow

CAUTION: The Restek Electronic Leak Detector may be used for determining trace amounts of hydrogenin a GC environment only. It is NOT designed for determining leaks of combustible gases. A combustiblegas detector should be used for determining combustible gas leaks under any condition.