Specifications
Parker Balston Zero Air Generators
|
|
|
| Maximum Zero Air Flow Rate: |
75-83NA
HPZA-3500 HPZA-7000 HPZA-18000 HPZA-30000 |
1 lpm
3.5 lpm 7 lpm 18 lpm 30 lpm |
| Outlet Hydrocarbon Concentration (as methane): |
75-83NA HPZA-30000 Other Models |
< 0.1 ppm < 0.1 ppm < .05 ppm |
| Minimum/Maximum Inlet Air Pressure: |
40 psig/125 psig (276/862kPa) |
|
| Maximum Inlet Hydrocarbon Concentration (as methane): |
100 ppm |
|
| Pressure Drop at Maximum Flow Rate: |
4 psi (28kPa) differential |
|
| Maximum Inlet Air Temperature: |
78°F (25°C) |
|
| Inlet/Outlet Ports: |
1/4" NPT (female) |
|
| Start-up Time to Specified Hydrocarbon Concentration: |
45 minutes |
|
| Electrical Requirements: |
75-83NA Other Models |
120 VAC/60 Hz, 0.5 amps 120 VAC/60 Hz, 3.5 amps |
| Dimensions: |
75-83NA Other Models |
12"h x 10"w x 3"d (30cm x 25cm x 8cm) 16"h x 11"w x 13"d (42cm x 27cm x 34cm)
|
| Shipping Weight: |
75-83NA Other Models |
7 lbs. (3 kg) 41 lbs. (19 kg) |
| Model |
Number of FIDs* |
| 75-83NA |
Up to 3 |
| HPZA-3500 |
Up to 11 |
| HPZA-7000 |
Up to 23 |
| HPZA-18000 |
Up to 60 |
| HPZA-30000 |
Up to 100 |
*based on a 300cc/min. fuel air rate