<
Educational Materials
<
Software
Rtx®-Wax: Fast GC Columns
Narrow bore (0.10mm ID) columns are attractive alternatives to conventional-diameter capillary columns because they provide faster analysis times and higher resolving power. As column ID decreases, column efficiency (plates/meter) greatly increases. For instance, a 0.18mm ID column (5,150 plates/meter) is much more efficient than a 0.25mm ID column (2,500 plates/meter). Therefore, resolution can be achieved with a shorter column, which decreases the analysis time. When switching from a 0.25mm ID column to a 0.10mm ID column (8,500 plates/meter), the improvement in column efficiency is even more dramatic.
Typically, 0.18mm ID columns are used for fast GC analysis, and methods are easily converted. The 0.10mm ID columns require more research to switch methods to the smaller ID due to higher back pressures and lower column capacity.
The outer diameter of the 0.10mm and 0.18mm ID tubing is the same as 0.25mm ID tubing, which makes connections less complicated.
- Significantly reduces analysis time without sacrificing resolution.
- Highest column efficiencies, great for GC/MS.
- Excellent for comprehensive GC (GCxGC) as second dimension column.
(Crossbond® Carbowax® polyethylene glycol)