Handy Septum Size Chart

Handy Septum Size Chart

Instrument Septum Diameter      Instrument Septum Diameter
Agilent Pye/Unicam
5880A, 5890, 6850,
6890, 7890, PTV
11mm All Models 7mm
5700, 5880 9.5/10mm Shimadzu
On-Column Injection 5mm All Models Plug
Thermo Scientific SRI
TRACE GC 17mm All Models Plug
GCQ w/TRACE, PTV 17mm Tracor
8000 Series 17mm 540 11.5mm
Finnigan (TMQ) 550, 560 9.5mm
GC 9001 9.5mm 220, 222 12.7mm
GCQ 9.5mm Varian
QCQ 9.5mm Injector Type
TRACE 2000 9.5mm Packed Column 9.5/10mm
Gow-Mac 1078/1079 10/11mm
6890 Series 11mm 1177 9mm
All Other Models 9.5mm 1075/1077 11mm
PerkinElmer
Sigma Series 11mm
900, 990 11mm
8000 Series 11mm
Auto SYS, Auto SYS XL 11mm

Tips for Handling Septa

All septa, regardless of composition, puncturability, or resistance to thermal degradation will fail if they are mishandled. Overtightening a septum nut invariably will reduce septum lifetime by increasing coring/splitting. All septa contain volatile materials (e.g., phthalates) that are released when the septum is heated (septum bleed). Because most GCs are equipped with a septum purge, septum bleed generally will disappear within 30 minutes after installing a new septum and exposing it to normal injector temperatures. All Restek septa are preconditioned and are ready to use without delay.