Specifications for Click-On Inline Super-Clean Purification Gas Traps
Click-On Traps measure: 8 1/2" x 1 1/4" (21.6 x 3.2cm)
| -----Capacity------ | ||||||||
| Type of Filter | Gas Quality at Outlet | Maximum Pressure | Maximum Flow (L/min.) | Use for: | H20 (g) | O2 (mL) | Hydrocarbons (g) (as n-butane) |
Estimated Lifetime (years) |
| Moisture cat.# 22467 |
>99.9999 | 11 bar 160 psi |
25 | Inert carrier gas, helium, air, H2 | 15 | NA | NA | >3 |
| Oxygen cat.# 22468 |
>99.9999 | 11 bar 160 psi |
25 | Inert carrier gas | NA | 2,000 | NA | >3 |
| Hydrocarbon cat.# 22466 |
>99.9999 | 11 bar 160 psi |
25 | Inert carrier gas, helium, air, H2 | NA | NA | 24 | >3 |
| Fuel Gas1 cat.# 22465 |
>99.9999 | 11 bar 160 psi |
25 | Inert carrier gas, helium, air, H2 | 7 | NA | 12 | >2 |
| Triple2 cat.# 22464 |
>99.9999 | 11 bar 160 psi |
25 | Inert carrier gas | 4 | 1,000 | 8 | >2 |
1 Removes hydrocarbons, moisture.
2 Removes hydrocarbons, moisture, oxygen.
NOTE:
Super-Clean Gas Filters are recommended for purifying non-corrosive gases with low concentrations of contaminants. The maximum concentration of oxygen in the incoming gas stream for oxygen purifiers is 0.5%.
Did you know?
Trap replacement depends on the quality of the incoming gas. Use the double connector and install an indicating cartridge after a trap to indicate when the trap should be replaced.
To prevent settling of desiccant, mount vertically, not horizontally.