delim
delim
Specifications for Parker Balston® Nitrogen Gas Generators for LC/MS

Specifications for Parker Balston® Nitrogen Gas Generators for LC/MS

NitroFlow Lab N2-04 N2-14 or
N2-14A
N2-22 or
N2-22A
N2-35 or
N2-35A
Maximum Nitrogen Flow Rate: 32 lpm 8 lpm 78scfh at 95% purity N2-22: 44 lpm
N2-22A: 29 lpm
44 lpm
Nitrogen Purity: 99.50% 99% 95.0%–99.5% 99% 99%
Min/Max Inlet Pressure: N/A 60 psig/145 psig 60 psig/145 psig 60 psig/145 psig 60 psig/145 psig
Electrical Requirements: 120VAC/60Hz None N2-14: None
N2-14A: 120VAC/60Hz
N2-22: None
N2-22A: 120VAC/60Hz
N2-35: None
N2-35A: 120VAC/60Hz
Dimensions: 27.6"h x 35.4"w x 12.2"d (70cm x 90cm x 31cm) 11"h x 13"w x 16"d (27cm x 34cm x 41cm) 50"h x 16"w x 16"d (127cm x 41cm x 41cm) 50"h x 16"w x 16"d (127cm x 41cm x 41cm) 50"h x 16"w x 16"d (127cm x 41cm x 41cm)
Shipping Weight: 205 lbs. (93 kg) 43 lbs. (20 kg) N2-14: 75 lbs. (34 kg)
N2-14A: 80 lbs. (36 kg)
N2-22: 101 lbs. (46 kg)
N2-22A: 106 lbs. (48 kg)
N2-35: 115 lbs. (52 kg)
N2-35A: 119 lbs. (54 kg)