Restek Capillary Column Installation Guide - Section IV: Column Troubleshooting & Return Procedure

Restek Capillary Column Installation Guide - Section IV: Column Troubleshooting & Return Procedure

  1. Pinpointing Performance Problems
  2. Contacting Restek for Technical Service
  3. Return Authorization Procedure

Many problems associated with capillary chromatography are due to sample contamination or instrument variables. Our technical service chemists may suggest some troubleshooting procedures to isolate these variables. Preliminary troubleshooting efforts by you help us identify the cause of the problem faster and can save you valuable down time.



A. Please follow these steps to pinpoint performance problems:


Isolate the Column From Instrument Variables:
The first step is to isolate the column from instrument variables. To do this, simply install a new column into the same GC without changing any GC related variables. We highly recommend keeping a spare column on hand at all times, but if you do not have a spare column, then install the questionable column in another GC that is working properly. If the problem does not seem to follow the column, (i.e. the new column does not work in the same GC, or the old column works fine in another GC), then we suggest performing routine maintenance on the injection port and detector as described in Section II of this installation guide. If the problem seems to follow the column, then it is time to perform routine maintenance on the column as described in Section III of this installation guide.

Begin Diagramming and Documenting
Document the procedures in chronological order listing times, dates, and important instrument parameters. Label all troubleshooting chromatograms. These steps help to inform anyone else that may be working on the system of the troubleshooting procedures that have been completed.

Start with a simple instrument diagram and try switching column A to detector B and vice versa. If the problem moves to detector B, then the problem is most likely occurring in the injector.



If the problem stays on Detector A when the column outlet is switched, then suspect a detector problem. Begin isolating detector problems by switching hardware, cables, electrometers, integrators, or any suspect part in the pathway. If the problem goes away from the A side, then that detector is most likely the cause and should be replaced.



B. If you are unable to pinpoint the problem:

Contact our technical service group at 800-356-1688 (ext. 4). Save all chromatograms and notes on any operating parameters.

Provide Restek's Technical Service Chemists with This Information:
  1. The column's catalog and serial numbers.
  2. Where and when the column was purchased and the invoice or purchase order number.
  3. Instrument parameters:
    1. instrument manufacturer and model number
    2. injection mode
    3. attenuation and range setting
    4. oven temperature profile
    5. injector and detector temperatures
    6. detector type
  4. Sample parameters:
    1. sample size, list of sample components, and concentrations
    2. description of the sample components that work and those that are problematic
    3. derivatization agents, inorganics, or high molecular weight residue which may be present
    4. number of injections made with the sample on this particular column and the length of time the column has been in service
  5. Concise description of the problem:
    1. observed symptoms
    2. variables affecting the symptoms
    3. steps taken to solve the problem
    4. if the problem has been observed before

Before you call, obtain chromatograms generated when the column was working properly and chromatograms that illustrate the problem you are having.



C. Return Authorization Procedure:


Procedure for Returning Capillary GC Columns
If a phone call does not resolve the problem satisfactorily, you can return the column after you obtain a return material authorization number (RMA#) by following our return instructions here.

Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us