Featured Application: Phosphatidylethanol (PEth) in Blood on Raptor FluoroPhenyl

High-Throughput LC-MS/MS Analysis of Phosphatidylethanol in Whole Blood


  • Simple protein precipitation sample prep streamlines workflow.
  • Fast 3.5-minute gradient cycle time supports faster sample throughput.
  • Easy-to-prepare mobile phases are quick to make consistently.

Phosphatidylethanol (PEth) is a group of phospholipids formed through an enzymatic reaction between ethanol and phosphatidylcholine on cell membranes. Among the multiple homologues of PEth, PEth-16:0/18:1 is the predominant molecule extracted from human erythrocytes. Analysis of phosphatidylethanol 16:0/18.1 is an important biomarker of alcohol consumption, and it can be measured in whole blood with a detection window of up to 3-4 weeks. Previously, PEth was considered a biomarker for high and sustained alcohol consumption, but with the application of highly sensitive LC-MS/MS techniques, it is now possible to use PEth concentration in blood to differentiate chronic drinking from social drinking or as a marker of absolute abstinence.

The LC-MS/MS method shown here was developed using a Raptor FluoroPhenyl column because of the column’s unique selectivity and retention mechanisms. Specific and sensitive analysis of phosphatidylethanol 16:0/18:1 in whole blood was achieved using a simple protein precipitation sample preparation procedure and a fast 3.5-minute LC cycle time. The high-quality chromatographic results that were achieved in a short analysis time make this approach ideal for labs needing a low-cost, high-throughput PEth method for monitoring alcohol consumption.


PeakstR (min)Conc.Precursor IonProduct Ion
1.Phosphatidylethanol-d5 (PEth-d5) 16:0/18:11.580.4 μM706.6281.3
2.Phosphatidylethanol (PEth) 16:0/18:11.580.5 μM701.5255.3
Phosphatidylethanol 16:0/18:1 in Human Whole Blood on Raptor FluoroPhenyl
LC_CF0733
ColumnRaptor FluoroPhenyl (cat.# 9319A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:40 °C
Sample
Conc.:0.5 μM in blood
Inj. Vol.:2 µL
Mobile Phase
A:Water, 5 mM ammonium acetate
B:9:1 Methanol:2-propanol
Time (min)Flow (mL/min)%A%B
0.000.53070
1.000.52080
2.000.52080
2.500.50100
2.510.53070
3.500.53070
DetectorMS/MS
Ion Mode:ESI-
Mode:MRM
InstrumentUHPLC
NotesSample Preparation Procedure (Provided by Redhot Diagnostics):
The fortified blood sample (50 µL) was mixed with 50 µL of internal standard (0.4 µM PEth-d5 in 2-propanol) and 150 µL of 4:1 2-propanol:tetrahydrofuran. The mixture was vortexed for 20 seconds at 3000 rpm and centrifuged for 10 minutes at 4300 rpm. The supernatant (2 µL) was injected for analysis.

Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us