share:
article_image

Analyze Haloacetic Acids in Under 13 Minutes with Rtx-CLPesticides Columns

By Jason Thomas, Environmental Innovations Chemist
  • Increase sample throughput with fast analysis times.
  • Complete separation of all target HAAs; two elution order changes.
  • Excellent for other GC/ECD methods—fewer column changes means less downtime.
Fully resolve HAAs under high throughput conditions.

Modern water treatment technologies provide safe, reliable drinking water and have substantially curbed outbreaks of water-borne diseases such as typhoid and cholera. However, the use of disinfectants in water treatment facilities can also lead to adverse health effects caused by the formation of disinfection byproducts (DBPs). DBPs are created by the reaction of the disinfection agent with naturally occurring organic matter and inorganic salts. Haloacetic acids (HAAs), for example, are a type of DBP that can form when chlorine is used as a disinfectant. DBP levels must be monitored in potable water supplies to ensure that the maximum allowable levels are not exceeded. Testing for HAA compounds in drinking water is usually done according to EPA Method 552.2; however, labs using this method struggle with low sample throughput due to long analytical run times.

Sample throughput can be significantly increased by choosing phases with a unique selectivity for HAAs. Using Rtx-CLPesticides and Rtx-CLPesticides2 columns all target HAAs were resolved in less than 13 minutes (Figure 1). Two elution order changes were observed between the columns: methyl trichloroacetate/1,2,3-trichloropropane and methyl bromodichloroacetate/methyl dibromoacetate. Notably, pairs which are extremely difficult to separate under method conditions, such as methyl dichloroacetate/dalapon methyl ester and methyl chlorodibromoacetate/methyl 2,3-dibromopropionate, were easily resolved on these columns.

In addition to performing well for haloacetic acid analysis, Rtx-CLPesticides and Rtx-CLPesticides2 columns give excellent results for other GC/ECD methods, including methods for chlorinated pesticides, chlorinated herbicides, and PCBs. Labs can increase HAA sample throughput and reduce the downtime spent changing columns between methods by switching to Rtx-CLPesticides and Rtx-CLPesticides2 columns.

For analysis of chlorinated pesticides, chlorinated herbicides, and PCBs, click here.

Figure 1  Increase sample throughput for haloacetic acids with Rtx-CLPesticides/Rtx-CLPesticides2 columns.

PeaksConc.
(µg/mL)
1.Methyl monochloroacetate1.2
2.Methyl monobromoacetate0.8
3.Methyl dichloroacetate1.2
4.Dalapon methyl ester2
5.Methyl trichloroacetate0.4
6.1,2,3-Trichloropropane (IS)4
PeaksConc.
(µg/mL)
7.Methyl bromo(chloro)acetate0.8
8.Methyl bromodichloroacetate0.8
9.Methyl dibromoacetate0.4
10.Methyl chlorodibromoacetate2
11.Methyl 2,3-dibromopropionate (SS)2
12.Methyl tribromoacetate4
Haloacetic Acids & Dalapon by EPA Method 552.2 on Rtx-CLPesticides & Rtx-CLPesticides2
GC_EV01006
ColumnsRtx-CLPesticides2 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324)
and Rtx-CLPesticides 30 m, 0.32 mm ID, 0.32 µm (cat.# 11141)
using Rxi guard column 5 m, 0.32 mm ID (cat.# 10039)
with deactivated universal “Y” Press-Tight connector (cat.# 20405-261)
SampleHaloacetic acid methyl ester mix #2 (cat.# 31647)
Dalapon methyl ester (cat.# 32057)
Methyl-2,3-dibromopropionate (cat.# 31656)
1,2,3-Trichloropropane (cat.# 31648)
Diluent:Methyl tert-butyl ether (MTBE)
Injection
Inj. Vol.:1.0 µL splitless (hold 0.75 min)
Liner:Cyclo double taper (4 mm) (cat.# 20896)
Inj. Temp.:250 °C
Oven
Oven Temp.:35 °C (hold 4 min) to 250 °C at 15 °C/min (hold 5 min)
Carrier GasHe, constant flow
Linear Velocity:25 cm/sec
DetectorMicro-ECD @ 300 °C
NotesThis chromatogram was obtained using an Agilent micro-ECD. To obtain comparable results, you will need to employ a micro-ECD in addition to confirmational dual columns connected to a 5-meter guard column using a “Y” Press-Tight connector.

Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us