Rapid Screening Method for Carbamates in Orange Oil

Using an Ultra Carbamate HPLC Column


By Julie Kowalski, PhD., Innovations Chemist
  • Fast analysis times, for increased sample throughput.
  • Simple methodology saves time—no sample preparation.
  • Accurate mass identification, for definitive results.

Concern over the presence of pesticides in food products, particularly citrus, is growing, resulting in an increasing number of countries regulating insecticides such as carbamates. EPA Method 531.1 describes a method for the analysis of carbamates in water, but not in other commodities. Matrices like citrus oil contain numerous interferences and often require time-consuming sample preparation. However, the method described here requires no sample preparation and provides fast analysis times, significantly increasing sample throughput.

Fast analysis times and no sample preparation allow faster sample throughput.

Carbamates are most easily determined via HPLC analysis because derivatization is required for GC analysis. The rapid screening method shown here uses the Ultra Carbamate HPLC column, which is designed specifically for analyzing carbamates and is compatible with both traditional detectors and mass spectrometry. This column works well with mass spectrometry amenable buffers and allows an initial mobile phase composition of 20% organic, which promotes complete ionization at the electrospray source.

Orange oil was spiked at 10ppm with a carbamate mix and analyzed (Figures 1-2). The monoisotopic masses and retention times were compared to an injected standard and found to match closely (Table I). The high mass accuracy of the Leco Unique TOF- MS allowed positive analyte identification, even in a complex mixture containing compounds with the same nominal mass (within 1 amu) as the target carbamate. By using the Ultra Carbamate column in conjunction with the Leco Unique TOF- MS, we were able to develop a quick, easy, and accurate screening method for carbamates in a complex matrix such as orange oil.


Figure 1  Reference standard carbamates resolve quickly on an Ultra Carbamate HPLC column. (extracted ion chromatograms)

Peaks
1.Aldicarb sulfone
2.Aldicarb sulfoxide
3.Oxamyl
4.Methomyl
5.3-Hydroxycarbofuran
Peaks
6.Aldicarb
7.Propoxur
8.Carbofurane
9.Carbaryl
10.Methiocarb
11.BDMC (IS)
Carbamates on Ultra Carbamate
LC_FF0473
ColumnUltra Carbamate (cat.# 9177352)
Dimensions:50 mm x 2.1 mm ID
Particle Size:3 µm
Pore Size:100 Å
Temp.:ambient
Sample531.1 Carbamate Pesticide Calibration Mixture (cat.# 32273)
4-bromo-3,5-dimethylphenyl-N-methylcarbamate (BDMC) (cat.# 32274)
Diluent:methanol
Conc.:50 µg/mL
Inj. Vol.:1 µL
Mobile Phase
A:2 mM ammonium acetate:methanol (v/v, 90/10)
B:2 mM ammonium acetate:methanol (v/v, 10/90)
Time (min)%A%B
0.008020
200100
250100
Flow:0.2 mL/min
DetectorLECO Unique® TOFMS
Run Length:25 min
Ionization Source Type:high flow ESI
Ion Mode:positive
Desolvation Temp.:130 °C
Nebulizing Pressure:100 kPa
Desolvation Gas (N2):4 L/min
Interface Temp.:120 °C
Nozzle Voltage:62 V
Capillary Voltage:2.75 kV
InstrumentAgilent 1100
AcknowledgementLECO Corporation

Figure 2  Positive identification of carbamates in orange oil injected with no sample preparation. (extracted ion chromatograms)

Peaks
1.Aldicarb sulfone
2.Aldicarb sulfoxide
3.Oxamyl
4.Methomyl
5.3-Hydroxycarbofuran
6.Aldicarb
7.Propoxur
8.Carbofuran
9.Carbaryl
10.Methiocarb
11.BDMC (IS)
Carbamates in Orange Oil on Ultra Carbamate
LC_FF0472
ColumnUltra Carbamate (cat.# 9177352)
Dimensions:50 mm x 2.1 mm ID
Particle Size:3 µm
Pore Size:100 Å
Temp.:ambient
Sample531.1 Carbamate Pesticide Calibration Mixture (cat.# 32273)
4-bromo-3,5-dimethylphenyl-N-methylcarbamate (BDMC) (cat.# 32274)
Diluent:methanol
Conc.:10 ppm spiked into unprocessed orange oil
Inj. Vol.:3 µL
Mobile Phase
A:2 mM ammonium acetate:methanol, 90:10
B:2 mM ammonium acetate:methanol, 10:90
Time (min)%B
0.0020
20100
25100
Flow:0.2 mL/min
DetectorLECO Unique® TOFMS
Ionization Source Type:high flow ESI
Ion Mode:positive
Nebulizing Pressure:100 kPa
Desolvation Gas (N2):4 L/min
Interface Temp.:120 °C
Nozzle Voltage:62 V
Capillary Voltage:2.75 kV

Table I  Carbamates were positively identified in matrix using both retention time and mass.

Calculated ion monoisotopic massStandard ion monoisotopic massStandard Retention Time (min.)Orange oil ion monoisotopic massOrange oil Retention Time (min.)
aldicarb sulfone [M+H]+223.075223.0993.81223.1423.67
aldicarb sulfoxide[M+H]+207.080207.1034.31207.1224.09
oxamyl[M+NH4]+237.102237.0854.97237.1104.41
methomyl[M+H]+163.054163.0745.84163.0865.36
3-hydroxycarbofuran[M+H]+238.108238.1218.32238.1287.73
aldicarb[M+H]+191.085191.0728
116.0751*
11.92116.052*11.53
propoxur[M+H]+210.113210.15213.53210.15313.14
carbofuran[M+H]+222.113222.14013.98222.12013.66
carbaryl[M+H]+202.087202.08415.48202.10115.17
methiocarb[M+H]+226.090226.09719.22226.06019.12
BDMC[M+H]+258.013258.04219.89258.00519.84

* m/z 116.052 is a fragment ion with higher intensity than the [M+H]+ ion and was used for identification in orange oil

RELATED SEARCHES

Carbamates

,

ultra carbamate

,

pesticide residue

,

epa 531.1

,

epa method 531.1

,

orange

,

orange oil

,

citrus oil

Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us