share:
article_image

Prepare Olive Oil Samples for Pesticide Residue Analysis in Half the Time With a Fraction of the Solvent Using dSPE

By Michelle Misselwitz1, Julie Kowalski1, Mark Crawford2, Michael Halvorson2, and Joan M. Stevens2
1. Restek Corporation, 2. Gilson, Inc.

Simplify and speed up sample preparation with Resprep dSPE tubes! Here we show the extraction and cleanup of pesticide residues from olive oil samples—twice as fast as GPC, with only a fraction of the solvent required for conventional SPE.

Increase sample throughput with a quick, easy sample preparation method, while protecting your column from matrix contamination.

Olive oil is considered a healthy fat source and is a staple in many recommended diets. However, concerns about potentially negative health effects associated with pesticide residues have increased consumer interest in testing. While organophosphorus pesticides are currently used in olive orchards to control pests, organochlorine pesticides are still tested, even though they are no longer in commercial use, because they are persistent organic pollutants. There are several existing methods for measuring pesticide residues in olive oil, all of which involve sample extraction and cleanup [1]. The common goal of these methods is to remove lipids that are harmful to the analytical system [2]. Efficient sample cleanup procedures are critical to maximizing sample throughput and minimizing labor and material costs. Here we demonstrate the efficiency of a dSPE cleanup procedure, as well as the capabilities of a method-specific analytical column.

Simple Procedure Uses Half the Time and Minimal Solvent

Sample extraction and cleanup can be accomplished with gel permeation chromatography (GPC), solid phase extraction (SPE), or dispersive solid phase extraction (dSPE) methods. However the dSPE method shown here is much less expensive than GPC (which requires specialized equipment) and uses substantially less solvent than comparable GPC or SPE methods (Table I).3 The method is simple to use and allows sample extraction and cleanup to be accomplished in half the time of other techniques (Table II).

Extraction and dSPE Cleanup for Pesticide Residues in Olive Oil
Test sample: A 1.5 mL sample of commercially obtained virgin olive oil was spiked with a standard organochlorine pesticide mix. The spiked sample was processed as follows:
    1. Dilute with 1.5 mL hexane.
    2. Add 6 mL of acetonitrile (ACN).
    3. Mix for 30 minutes on a shaker.
    4. Allow layers to separate (approximately 20 minutes), then collect the top (ACN) layer.
    5. Repeat the liquid-liquid extraction (steps 2–4) and combine both ACN extract layers.
    6. Place 1 mL of the combined ACN extract in a 1.5 mL tube containing 150 mg magnesium
        sulfate and 50 mg PSA.
    7. Shake the tube for 2 minutes.
    8. Centrifuge at 3,000 U/min for approximately 5 minutes.
    9. Remove the top layer and inject directly into the gas chromatograph system.

Extracts were analyzed using the Rtx-CLPesticides2 column (Figure 1). The Rtx-CLPesticides2 column is a method-specific column that resolves all chlorinated pesticide residues tested. Recoveries of 70%–80% were obtained, levels comparable to conventional SPE—without the necessity of vacuum manifolds or high-pressure systems. The GPC method attained recoveries of > 95%. However this method requires large amounts of solvent and takes over twice as long as other methods.

The dSPE method shown here is an efficient, cost-effective way to clean up chlorinated pesticide residues in olive oil. With good recoveries and minimal matrix interference, it is an easy way to reduce solvent usage, compared to conventional SPE, and is more cost-effective than GPC.

Table I: The Resprep dSPE method uses 42% and 89% less solvent than SPE and GPC methods respectively.


The Restek dSPE method uses 42% and 89% less solvent than SPE and GPC methods respectively.

Table II: Cut extraction/cleanup time by 50% using a Resprep dSPE method.


Cut extraction/cleanup time by 50% using a Resprep dSPE method.

Figure 1:  Chlorinated pesticide residues in olive oil are easily separated on the Rtx-CLPesticides2 column.


PeaksQuant. ionQual. ion 1Qual. ion 2
1.α-BHC*219181109
2.γ-BHC219181109
3.β-BHC219181109
4.δ-BHC219181109
5.Heptachlor272237100
6.Aldrin263293220
7.Heptachlor epoxide26323781
8.δ-Chlordane27223765
9.α-Chlordane27223765
10.Endosulfan I195207241
11.4,4'-DDE246318176
12.Dieldrin79263277
13.Endrin26328181
14.4,4'-DDD235165199
15.Endosulfan II195207-
16.4,4'-DDT235165199
17.Endrin aldehyde67250345
18.Endosulfan sulfate272229239
19.Methoxychlor227274-
20.Endrin ketone67317281
* α-BHC not detected due to low recovery during sample preparation.
Chlorinated Pesticide Residues in Olive Oil Rtx-CLPesticides2
GC_FF01043
ColumnRtx-CLPesticides2, 30 m, 0.25 mm ID, 0.20 µm (cat.# 11323)
SampleOlive oil spiked with Organochlorine Pesticide Mix AB # 3 (cat.# 32415)
Conc.:10 µg/mL
Injection
Inj. Vol.:1 µL splitless (hold 0.5 min)
Liner:Single Taper w/Wool (cat.# 22286-200.1)
Inj. Temp.:225 °C
Oven
Oven Temp.:140 °C (hold 0.5 min) to 268 °C at 20 °C/min to 290 °C at 3 °C/min to 330 °C at 20 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1 mL/min
DetectorMS
Mode:SIM
Transfer Line Temp.:320 °C
Ionization Mode:EI
NotesExtraction and dSPE Cleanup for Pesticide Residues in Olive Oil

Test sample: A 1.5 mL sample of commercially obtained virgin olive oil was spiked with a standard organochlorine pesticide mix. The spiked sample was processed as follows:
1. Dilute with 1.5 mL hexane.
2. Add 6 mL of acetonitrile (ACN).
3. Mix for 30 minutes on a shaker.
4. Allow layers to separate (approximately 20 minutes), then collect the top (ACN) layer.
5. Repeat the liquid-liquid extraction (steps 2–4) and combine both ACN extract layers.
6. Place 1 mL of the combined ACN extract in a 1.5 mL tube containing 150 mg magnesium sulfate and 50 mg PSA.
7. Shake the tube for 2 minutes.
8. Centrifuge at 3,000 U/min for approximately 5 minutes.
9. Remove the top layer and inject directly into the gas chromatograph system.

References

[1] C. Lentza-Rizos, E.J. Avramides, Rev. Environ. Contam. Toxicol. 141 (1995) 111.
[2] S. Cunha, S. Lehotay, K. Mastovska, J. Sep. Sci. 30 (2007) 620.
[3] M. Crawford, M. Halvorson, J. Stevens, The Examination and Automation of GPC, SPE and QuEChERS Utilized in Extracting Pesticides from Olive Oil. HPLC 2008 poster presentation.

RELATED SEARCHES

Pesticide Residue

,

sample preparation

,

SPE

,

dSPE

Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us