Guide to GC Column Selection and Optimizing Separations
- Learn how to choose the right column the first time.
- Optimize separations for the best balance of resolution and speed.
- Troubleshoot quickly and effectively based on chromatographic symptoms.
You can improve lab productivity by assuring that speed and resolution are optimized. One of the best ways to do this is to use the resolution equation (Figure 1) as the key to controlling your separations. This fundamental equation helps you choose the best column stationary phase, length, inner diameter (ID), and film thickness for your specific applications. Once you understand the basics of how resolution is related to column characteristics, optimizing your analysis for both separation and speed becomes easier. This GC column selection guide discusses the basics of separation and teaches you how to choose the right GC column!
Resolution is the goal of every chromatographer, but how much resolution is enough? Practically speaking, we need enough retention to get sharp symmetrical peaks that are baseline resolved from each other, but not so much retention, that retention times are too long, and peaks start to broaden. To achieve this goal, we must consider the column and non-column factors that affect our “perfect separation.” Only then can we work towards selecting the right column and optimizing GC separations and analysis speed. Now, let’s consider the separation factor (α), retention factor (k), and efficiency (N) in turn and how they can help you select the right column and optimize your separation.
Figure 1: The resolution equation and factors that affect it.
Shortcut to Column Selection
- Look for application-specific stationary phases first; these columns are optimized for specific analyses and will provide the best resolution in the shortest time (Table III).
- If an application-specific column is not available, and you need to measure low concentrations or are using a mass spectrometer (MS), then choose an Rxi column. Rxi technology unites outstanding inertness, low bleed, and high reproducibility, resulting in high-performance GC columns that are ideal for trace analysis and MS work (Table II).
- For other methods, choose a general-purpose Rtx column (Table II).
Use Separation Factor (α) to Choose the Best Stationary Phase
Choosing the right stationary phase is the first step toward optimizing your GC separation. It is the most important decision you will make because the separation factor (α) has the greatest impact on resolution, and it is strongly affected by stationary phase polarity and selectivity.
Stationary phase polarity is determined by the type and amount of functional groups in the stationary phase. When choosing a column, consider the polarity of both the stationary phase and your target analytes. If the stationary phase and analyte polarities are similar, then the attractive forces are strong, and more retention will result. Greater retention often results in increased resolution. Stationary phase polarity strongly influences column selectivity and the separation factor, making it a useful consideration when selecting a column.
Stationary phase selectivity is defined by IUPAC as the extent to which other substances interfere with the determination of a given substance. Selectivity is directly related to stationary phase composition and how it interacts with target compounds through intermolecular forces (e.g., hydrogen bonding, dispersion, dipole-dipole interactions, and shape selectivity). As methyl groups in the stationary phase are replaced by different functionalities, such as phenyl or cyanopropyl pendant groups, compounds that are more soluble with those functional groups (e.g., aromatics or polar compounds, respectively) will inter- act more and be retained longer, often leading to better resolution and increased selectivity. In another example of the effect of stationary phase-analyte interactions, an Rtx-200 stationary phase is highly selective for analytes containing lone pair electrons, such as halogen, nitrogen, or carbonyl groups, due to interactions with the fluorine pendant group in this phase. Selectivity can be approximated using existing applications or retention indices (Table I), making these useful tools for comparing phases and deciding which is most appropriate for a specific analysis.
Due to their influence on the separation factor, polarity and selectivity are primary considerations when selecting a column. However, temperature limits must also be considered. In general, highly polar stationary phases have lower maximum operating temperatures, so choosing a column with the appropriate maximum operating temperature, as well as optimal polarity and selectivity for the type of compounds being analyzed, is crucial. Use Table II and Figure 2 to determine which general-purpose column is most appropriate based on the selectivity, polarity, and the temperature requirements of your analysis. See Table III for a list of specialty stationary phases designed for specific applications.
Tech Tip
In many cases, different GC oven temperature programs can change the elution order of sample analytes on the same column. Reconfirm elution orders if changing GC oven temperature programs.
Table I: Kovat’s retention indices for GC phases can be used to approximate selectivity.
Stationary Phase | Benzene | Butanol | Pentanone | Nitropropane |
100% Dimethyl polysiloxane | 651 | 651 | 667 | 705 |
5% Diphenyl/95% dimethyl polysiloxane | 667 | 667 | 689 | 743 |
20% Diphenyl/80% dimethyl polysiloxane | 711 | 704 | 740 | 820 |
6% Cyanopropylphenyl/94% dimethyl polysiloxane | 689 | 729 | 739 | 816 |
35% Diphenyl/65% dimethyl polysiloxane | 746 | 733 | 773 | 867 |
Trifluoropropylmethyl polysiloxane | 738 | 758 | 884 | 980 |
Phenyl methyl polysiloxane | 778 | 769 | 813 | 921 |
14% Cyanopropylphenyl/86% dimethyl polysiloxane | 721 | 778 | 784 | 881 |
65% Diphenyl/35% dimethyl polysiloxane | 794 | 779 | 825 | 938 |
50% Cyanopropylmethyl/50% phenylmethyl polysiloxane | 847 | 937 | 958 | 958 |
Polyethylene glycol | 963 | 1158 | 998 | 1230 |
STATIONARY PHASEFigure 2: Polarity scale of common stationary phases. |
Tech Tip
Any homologous series of compounds, that is, analytes from the same chemical class (e.g., all alcohols, all ketones, or all aldehydes, etc.) will elute in boiling point order on any stationary phase. However, when different compound classes are mixed together in one sample, intermolecular forces between the analytes and the stationary phase are the dominant separation mechanism, not boiling point.
Table II: Relative polarity and maximum temperature are important considerations when selecting a GC stationary phase.
Restek | Phase Composition (USP Nomenclature) | Restek’s Max Temps* | Agilent | Phenomenex |
Rxi-1HT, Rxi-1ms, Rtx-1 |
100% Dimethyl polysiloxane (G1, G2, G38) | 400 °C 350 °C |
HP-1/HP-1ms, DB-1/DB-1ms, VF-1ms, CP Sil 5 CB, Ultra 1, DB-1ht, HP-1ms UI, DB-1ms UI |
ZB-1, ZB-1MS, ZB-1HT Inferno |
Rxi-5HT, Rtx-5ms, Rxi-5ms, Rtx-5 |
5% Diphenyl/95% dimethyl polysiloxane (G27, G36) | 400 °C 350 °C |
HP-5/HP-5ms, DB-5, Ultra 2, DB-5ht, VF-5ht, CP-Sil 8 CB |
ZB-5, ZB-5HT Inferno, ZB-5ms |
Rxi-5Sil MS | 5% (1,4-bis(dimethylsiloxy) phenylene/95% dimethyl polysiloxane | 350 °C | DB-5ms UI, DB-5ms,VF-5ms |
ZB-5msi |
Rxi-XLB | Proprietary Phase | 360 °C | DB-XLB, VF-Xms | MR1, ZB-XLB |
Rtx-20 | 20% Diphenyl/80% dimethyl polysiloxane (G28, G32) | 320 °C | — | — |
Rtx-35 | 35% Diphenyl/65% dimethyl polysiloxane (G42) | 320 °C | HP-35, DB-35 | ZB-35 |
Rxi-35Sil MS | Proprietary Phase | 360 °C | DB-35ms, DB-35ms UI, VF-35ms | MR2 |
Rtx-50 | Phenyl methyl polysiloxane (G3) | 320 °C | — | — |
Rxi-17 | 50% Diphenyl/50% dimethyl polysiloxane | 320 °C | DB-17ms, VF-17ms, CP Sil 24 CB |
ZB-50 |
Rxi-17Sil MS | Proprietary Phase | 360 °C | DB-17ms, VF-17ms, CP Sil 24 CB |
ZB-50 |
Rtx-65 | 65% Diphenyl/35% dimethyl polysiloxane (G17) | 300 °C | — | — |
Rxi-624Sil MS | Proprietary Phase | 320 °C | DB-624 UI, VF-624ms, CP-Select 624 CB | ZB-624 |
Rtx-1301, Rtx-624 |
6% Cyanopropylphenyl/94% dimethyl polysiloxane (G43) | 280 °C 240 °C |
DB-1301, DB-624, CP-1301, VF-1301ms, VF-624ms |
ZB-624 |
Rtx-1701 | 14% Cyanopropylphenyl/86% dimethyl polysiloxane (G46) | 280 °C | DB-1701, VF-1701ms, CP Sil 19 CB, VF-1701 Pesticides, DB-1701R |
ZB-1701, ZB-1701P |
Rtx-200 | Trifluoropropyl methyl polysiloxane (G6) | 360 °C | DB-200, VF-200ms, DB-210 | — |
Rtx-200ms | Trifluoropropyl methyl polysiloxane (G6) | 340 °C | DB-200, VF-200ms, DB-210 | — |
Rtx-225 | 50% Cyanopropyl methyl/50% phenylmethyl polysiloxane (G7, G19) | 240 °C | DB-225ms, CP Sil 43 CB |
— |
Rtx-440 | Proprietary Phase | 340 °C | RESTEK INNOVATION | |
Rtx-2330 | 90% Biscyanopropyl/10% cyanopropylphenyl polysiloxane (G48) | 275 °C | VF-23ms | — |
Rt-2560 | Biscyanopropyl polysiloxane | 250 °C | HP-88, CP Sil 88 | — |
Rtx-Wax | Polyethylene glycol (G14, G15, G16, G20, G39) | 250 °C | CP-Wax 52 CB, DB-Wax, DB-WAX UI |
ZB-WAX |
Stabilwax | Polyethylene glycol (G14, G15, G16, G20, G39) | 260 °C | HP-INNOWax, VF-WaxMS |
ZB-WAXPlus |
* Maximum operating temperatures may vary with column film thickness.
Table III: Application-specific phases designed for particular analyses.
Restek | Applications | Agilent | Supelco | Macherey-Nagel | SGE | Phenomenex |
Rtx-Volatile Amine | Volatile amines | CP-VolAmine | — | — | — | — |
Rtx-5Amine | Amines | CP-Sil 8 CB | — | OPTIMA 5 Amine | — | — |
Rtx-35Amine | Amines | RESTEK INNOVATION |
||||
Stabilwax-DB | Amines | CAM, CP WAX 51 | Carbowax Amine | FS-CW 20 M-AM | — | — |
Stabilwax-DA | Free fatty acids | HP-FFAP, DB-FFAP, VF-DA, CP WAX58 CB, CP-FFAP CB |
Nukol | PERMABOND FFAP, OPTIMA FFAP, OPTIMA FFAP Plus |
BP-21 | ZB-FFAP |
Chiral Columns |
||||||
Rt-βDEXm, Rt-βDEXsm, Rt-βDEXse, Rt-βDEXsp, Rt-βDEXsa, Rt-βDEXcst, Rt-γDEXsa |
Chiral compounds | — | — | — | — | — |
Foods, Flavors, & Fragrances |
||||||
Rt-2560 | cis/trans FAMEs | HP-88 | SPB-2560 | — | — | — |
FAMEWAX | Marine oils | DB-FATWAX UI | Omegawax | — | — | — |
Rtx-65TG | Triglycerides | — | — | — | — | — |
Rxi-PAH | Polycyclic aromatic hydrocarbons (PAHs) | Agilent Select PAH | — | — | — | — |
Petroleum & Petrochemical |
||||||
Rxi-LAO | Linear alpha olefin impurities | RESTEK INNOVATION |
||||
Rt-Alumina BOND/CFC | Chlorinated fluorocarbons (CFCs) | RESTEK INNOVATION |
||||
Rt-Alumina BOND/MAPD | Trace analysis of methyl acetylene, propadiene, and acetylene | Select Al₂0₃ MAPD | ||||
Rtx-DHA | Detailed hydrocarbon analysis | HP-PONA, DB-Petro, CP Sil PONA CB | Petrocol DH | — | BP1PONA | — |
Rtx-2887, MXT-2887 |
Hydrocarbons (ASTM D2887) | DB-2887 | Petrocol 2887, Petrocol EX2887 |
— | — | — |
D3606 | Ethanol (ASTM D3606) | RESTEK INNOVATION |
||||
Rt-TCEP | Aromatics and oxygenates in gasoline | CP-TCEP | TCEP | — | — | — |
MXT-1HT SimDist | Simulated distillation | DB-HT-SimDis, CP-SimDist, CP-SimDist Ultimetal |
— | — | BPX1 | ZB-1XT SimDist |
Rtx-Biodiesel TG, MXT-Biodiesel TG |
Triglycerides in biodiesel | Biodiesel, Select Biodiesel |
— | OPTIMA Biodiesel | — | ZB-Bioethanol |
Clinical/Forensic |
||||||
Rtx-BAC Plus 1 | Blood alcohol testing | DB-ALC1 | — | — | — | ZB-BAC1 |
Rtx-BAC Plus 2 | Blood alcohol testing | DB-ALC2 | — | — | — | ZB-BAC2 |
Pharmaceutical |
||||||
Rxi-624Sil MS (G43) | Organic volatile impurities (USP <467>) | DB-624, VF-624ms, CP-Select 624 CB |
— | OPTIMA 624 LB | BP624 | ZB-624 |
Rtx-5 (G27) | Organic volatile impurities (USP <467>) | HP-5, DB-5, CP Sil 8 CB |
SPB-5 | OPTIMA 5 | BP5 | ZB-5 |
Stabilwax (G16) | Organic volatile impurities (USP <467>) | HP-INNOWax, CP Wax 52 CB, VF-WAX MS |
Supelcowax-10 | OPTIMA WAXplus | — | ZB-WAXplus |
Environmental |
||||||
Rxi-SVOCms | Semivolatiles, PAHs | DB-UI 8270D | ZB-SemiVolatiles | |||
Rxi-5Sil MS | Semivolatiles, PAHs | DB-5ms, DB-5msUI, VF-5ms, CP-Sil 8 CB |
SLB-5ms | OPTIMA 5MS Accent | BPX5 | ZB-5msi |
Rtx-VMS | Volatiles (EPA Methods 8260, 624, 524) | RESTEK INNOVATION |
||||
Rxi-624Sil MS | Volatiles (EPA Methods 624) | DB-624, VF-624ms, CP-Select 624 CB |
— | OPTIMA 624 LB | BP624 | ZB-624 |
Rtx-502.2 | Volatiles (EPA Methods 8010, 8020, 502.2, 601, 602) | DB-502.2 | VOCOL | — | — | — |
Rtx-Volatiles | Volatiles (EPA Methods 8010, 8020, 502.2, 601, 602) | — | VOCOL | — | — | — |
Rtx-VRX | Volatiles (EPA Methods 8010, 8020, 502.2, 601, 602) | DB-VRX | — | — | — | — |
Rtx-CLPesticides | Organochlorine pesticides | RESTEK INNOVATION |
||||
Rtx-CLPesticides2 | Organochlorine pesticides | RESTEK INNOVATION |
||||
Rtx-1614 | Brominated flame retardants | RESTEK INNOVATION |
||||
Rtx-PCB | Polychlorinated biphenyl (PCB) congeners | RESTEK INNOVATION |
||||
Rxi-XLB | Polychlorinated biphenyl (PCB) congeners | DB-XLB,VF-XMS | — | — | — | MR1, ZB-XLB |
Rtx-OPPesticides | Organophosphorus pesticides | RESTEK INNOVATION |
||||
Rtx-OPPesticides2 | Organophosphorus pesticides | RESTEK INNOVATION |
||||
Rtx-Dioxin2 | Dioxins and furans | RESTEK INNOVATION |
||||
Rtx-Mineral Oil | DIN EN ISO 9377-2 | Select Mineral Oil | — | — | — | — |
Select Column Film Thickness and Column ID Based on Retention Factor
Once you have chosen the stationary phase, you need to determine which column film thickness and inner diameter combination will give the retention factor (k) needed for optimal resolution and speed. Retention factor is sometimes referred to as “capacity factor,” which should not be confused with sample loading capacity.
The retention factor (k) of a column is based on the time an analyte spends in the stationary phase rela- tive to the time it spends in the carrier gas. As a general rule, the thicker the film and the smaller the inner diameter, the more an analyte will be retained. Note that as temperature increases k decreases, so at higher temperatures analytes stay in the carrier gas longer and are less retained.
In practice, if the value of k is too large, the peak will broaden, which can reduce resolution by causing peaks to overlap or coelute. Narrow, symmetrical peaks are important to maximizing resolution, so the goal is to select a column with a sufficient retention factor, such that resolution occurs, and peak shape does not suffer. Once the proper stationary phase is selected, column film thickness, column inner diam- eter, and elution temperature should be optimized to produce an acceptable retention factor.
Tech Tip
The sample loading capacity of the column must also be considered; if the mass of the target analyte exceeds the sample loading capacity of the column, loss of resolution, poor reproducibility, and fronting peaks will result. A larger ID column with thicker film is recommended for higher concentration samples, such as purity analysis, to minimize sample overload.
Film Thickness
Film thickness (μm) has a direct effect on both the retention of each sample component and the maximum operating temperature of the column. When analyzing extremely volatile compounds, a thick film column should be used to increase retention; more separation is achieved because the compounds spend more time in the stationary phase. If analyzing high molecular weight compounds, a thinner film column should be used as this reduces the length of time that the analytes stay in the column and minimizes phase bleed at higher elution temperatures. Use Figure 3 to select the best film thickness for your application. Note that as a general rule, the thicker the film, the lower the maximum temperature; exceeding the maximum temperature can result in column bleed and should be avoided.
FILM THICKNESSFigure 3: Characteristics and recommended applications based on film thickness. |
Tech Tip
Remember, when changing either film thickness and/or the temperature program, you must reconfirm peak identifications as elution order changes can occur.
Inner Diameter (ID)
Column ID does not have as great an effect on retention factor as film thickness does. However, when selecting column ID with retention factor (k) in mind, a general rule of thumb applies; smaller ID columns produce higher retention factors compared to larger ID columns. This is due to less available mobile phase (carrier gas) volume in the column. Because smaller ID columns produce higher k values, they are more suited towards complex sample analysis where a range of low to high molecular weight compounds may exist in the sample (Figure 4). Keep in mind that both ID and film thickness should be optimized together to produce the best resolution and peak shape.
INNER DIAMETERFigure 4: Characteristics and recommended applications based on column inner diameter. |
Tech Tip
When choosing column ID, the injection technique is also important because the column ID may need to be selected based on whether a split, splitless, direct, cool on-column injection, or other sample transfer method is being used. For example, 0.53 mm ID columns are ideal for cool on-column injections since the syringe needle (26 gauge) will fit into the large column ID. In addition, the detector and its optimal flow rate must be considered. Some MS detectors can only operate under column flow rates of up to 1.5 mL/min; therefore, a 0.53 mm ID column, which requires higher flows for proper chromatography, is not an option for MS work.
Phase Ratio (β)
The relationship between column inner diameter and stationary phase film thickness is expressed as phase ratio (β). If a good separation has been achieved on a larger diameter column, and a faster anal- ysis is desired, this can often be accomplished by reducing the inner diameter of the column without sacrificing, and sometimes even improving, separation efficiency. To maintain a similar compound elution pattern when narrowing column inner diameter, film thickness must also be changed. By choosing a column with a similar phase ratio, it will be easier to translate your application to the new column. Phase ratios for common column dimensions are given in Table IV. As shown here, an analyst wanting to decrease analysis time could switch from a 0.32 mm x 0.50 µm column (β = 160) to a 0.25 mm x 0.25 μm column (β = 250) and obtain a very similar separation upon proper method translation. Importantly, column inner diameter and stationary phase film thickness show a combined effect when it comes to sample loading capacity, which is decreased as column inner diameter and film thickness are reduced. It may be necessary to inject a lower sample amount in this case.
Table IV: Phase ratio (β)* values for common column dimensions. To maintain similar separations, choose columns with similar phase ratios when changing to a column with a different inner diameter or film thickness.
Column ID | Film Thickness (df) | ||||||
0.10 µm | 0.25 µm | 0.50 µm | 1.0 µm | 1.5 µm | 3.0 µm | 5.0 µm | |
0.18 mm | 450 | 180 | 90 | 45 | 30 | 15 | 9 |
0.25 mm | 625 | 250 | 125 | 63 | 42 | 21 | 13 |
0.32 mm | 800 | 320 | 160 | 80 | 53 | 27 | 16 |
0.53 mm | 1325 | 530 | 265 | 128 | 88 | 43 | 27 |
*Phase ratio (β) = radius/2df (Note: Convert variables to the same units prior to calculation.)
Consider Efficiency when Choosing Column Length, Column ID, and Carrier Gas
Column Length
Capillary GC columns are made in various lengths, typically 10, 15, 30, 60, and 105 meters, depending on the inner diameter. Longer columns provide more resolving power than shorter columns of the same inner diameter, but they also increase analysis time and should be used only for applications demanding the utmost in separation power. Column length should only be considered once the stationary phase has been determined. This is because the separation factor has the greatest effect on resolution and it is maximized through proper stationary phase choice for the compounds of interest. Doubling the column length (e.g., 30 m to 60 m) increases resolution by approximately 40%, while analysis time can be twice as long. In addition, longer columns cost more. Conversely, if a separation can be performed on a shorter column (e.g., 15 m versus 30 m), then both analysis time and column cost will be less. Figure 5 summarizes the characteristics and general application parameters for a range of typical column lengths.
LENGTHFigure 5: Characteristics and recommended applications based on column length. |
Inner Diameter (ID)
Compared to larger ID columns, smaller ID columns generate more plates per meter and sharper peaks, leading to better separation efficiencies. When more complex samples need to be analyzed, smaller ID columns can produce better separation of closely eluting peaks than larger ID columns. However, sample loading capacities are lower for smaller ID columns. Smaller ID columns, especially those at 0.18 mm and less, demand highly efficient injection techniques so that the column efficiency is not lost at the point of sample introduction. Column characteristics based on ID are presented in Table V.
Generally speaking, a 0.25 mm column will produce the most efficient sample analysis while simultaneously considering analysis time and sample loading capacity. For these reasons, in combination with its relatively low outlet flow, it is also the best column choice for GC-MS work.
Table V: General column characteristics based on ID.
Characteristic | Column Inner Diameter (mm) | |||||
0.10 | 0.15 | 0.18 | 0.25 | 0.32 | 0.53 | |
Nitrogen flow (mL/min) | 0.2 | 0.3 | 0.3 | 0.4 | 0.6 | 0.9 |
Helium flow (mL/min) | 0.6 | 0.8 | 1.0 | 1.4 | 1.8 | 3.0 |
Hydrogen flow (mL/min) | 0.7 | 1.1 | 1.8 | 1.8 | 2.3 | 3.7 |
Sample loading capacity (ng) | 2.5 | 10 | 20 | 50 | 125 | 500 |
Theoretical plates/meter | 11,000 | 7000 | 6000 | 4000 | 3000 | 2000 |
Note: Flows listed are for maximum efficiency. Sample loading capacities are estimates only. Actual sample loading capacity varies with film thickness and analyte.
Tech Tip
When changing carrier gas flow rates, you must reconfirm peak identifications as elution order changes can occur.
Carrier Gas Type and Linear Velocity
Carrier gas choice and linear velocity significantly affect column separation efficiency, which is best illustrated using van Deemter plots (Figure 6). The optimum linear velocity for each gas is at the lowest point on the curve, where plate height (H) is minimized, and efficiency is maximized. As seen in Figure 6, the optimum linear velocities differ among common carrier gases.
Nitrogen provides the best efficiency; however, the steepness of its van Deemter plot on each side of optimum means that small changes in linear velocity can result in large negative changes in efficiency. Compared to nitrogen, helium has a wider range for optimal linear velocity, but offers slightly less efficiency. In addition, because of its optimum velocity being faster, analysis times with helium are about half those when using nitrogen, and there is only a small sacrifice in efficiency when velocity changes slightly. Of the three common carrier gases, hydrogen has the flattest van Deemter curve, which results in the shortest analysis times and the widest range of average linear velocity over which high efficiency is obtained.
Regardless of the type of gas used, the carrier gas head pressure is constant during column temperature programming, whereas the average linear velocity decreases during the run. For constant pressure work then, the optimal linear velocity should be set for the most critical separations. More common today, electronic pneumatic control (EPC) of carrier gas allows for constant flow or even constant linear velocity, which helps maintain high efficiency throughout a temperature programmed run.
Another consideration for carrier gas type that is important, even if not directly related to column efficiency, is whether a mass spectrometer (MS) is used as a vacuum-outlet detector for GC. In almost all cases, helium is the carrier gas of choice, not only for its chromatographic efficiency but also because it is easier to pump than hydrogen. Hydrogen can be reactive in MS sources, leading to undesirable spectrum changes for some compounds. Nitrogen is typically not a carrier gas option for GC-MS, as it severely reduces sensitivity.
Figure 6: Operating carrier gas at the optimum linear velocity will maximize efficiency at a given temperature. Red circles indicate optimum linear velocities for each carrier gas.
GC Troubleshooting Tips
Basic Steps
Follow these basic troubleshooting steps to isolate problems related to the sample, injector, detector, and column. Check the obvious explanations first and change only one thing at a time until you identify and resolve the problem.
Check the Obvious:
|
|
|
Identify the Cause:
- Define the problem clearly; for example, “Over the last 4 days, only the phenols in my sample have been tailing.”
- Review sample and maintenance records to identify trends in the data or problem indicators, such as area counts decreasing over time or injector maintenance not being performed as scheduled.
- Use a logical sequence of steps to isolate possible causes.
Document Work and Verify System Performance:
- Document all troubleshooting steps and results; this may help you identify and solve the next problem faster.
- Always inject a test mix and compare it to previous data to ensure restored performance.
Example Troubleshooting Sequence
An analyst observed that no peaks appeared during a GC-FID analysis. The flowchart below shows a logical progression of steps that can be used to identify the cause and correct the problem.
Symptoms and Solutions
Good chromatography is critical to obtaining accurate, reproducible results. Coelutions, asymmetric peaks, baseline noise, and other issues are common challenges in the GC laboratory. These analytical problems and others can be overcome by troubleshooting your separations using the tips below.
Poor Resolution
Cause | Solutions | |
Non selective stationary phase |
|
|
Poor efficiency |
|
|
Sample overload |
|
|
Incorrect analytical conditions used |
|
Poor Retention Time Reproducibility
Cause | Solutions | |
Leaks |
|
|
Analyte adsorption |
|
|
Resolution/integration issues |
|
|
Incorrect column/oven temperature program |
|
|
Incorrect or variable carrier gas flow rate/linear velocity |
|
|
Poor control of oven temperature programming |
|
|
Incorrect oven equilibration time |
|
|
If manual injection, delay between pushing start and actual injection |
|
Fronting Peaks
Cause | Solutions | |
Incompatible stationary phase |
|
|
Column overloading |
|
Tailing Peaks
Cause | Solutions | |
Adsorption due to surface activity or contamination |
|
|
Adsorption due to chemical composition of compound |
|
|
Leak in system |
|
|
Installation issues |
|
Split Peaks
Cause | Solutions | |
Mismatched solvent/stationary phase polarity |
|
|
Incomplete vaporization |
|
|
Sample loading capacity exceeded |
|
|
Fast autosampler injection into open liner |
|
Carryover/Ghost Peaks
Cause | Solutions | |
Contaminated syringe or rinse solvent |
|
|
Backflash (sample volume exceeds liner volume) |
|
|
Last analysis ended too soon |
|
High Bleed
Cause | Solutions | |
Improper column conditioning |
|
|
Contamination |
|
|
Leak in system and oxidation of stationary phase |
|
Unstable Baseline (Spiking, Noise, Drift)
Cause | Solutions | |
Carrier gas leak or contamination |
|
|
Injector or detector contamination |
|
|
Column contamination or stationary phase bleed |
|
|
Septum coring/bleed |
|
|
Loose cable or circuit board connections |
|
|
Variable carrier gas or detector gas flows |
|
|
Detector not ready |
|
Response Variation
Cause | Solutions | |
Sample issues |
|
|
Syringe problems |
|
|
Electronics |
|
|
Dirty or damaged detector |
|
|
Flow/temperature settings wrong or variable |
|
|
Adsorption/reactivity |
|
|
Leaks |
|
|
Change in sample introduction/injection method |
|
No Peaks
Cause | Solutions | |
Injection problems |
|
|
Broken column |
|
|
Column installed into wrong injector or detector |
|
|
Detector problems |
|
Broad Peaks
Cause | Solutions | |
High dead volume |
|
|
Low flow rates |
|
|
Slow GC oven program |
|
|
Poor analyte/solvent focusing |
|
|
Column film is too thick |
|
|
Sample carryove |
|
Propel Method Development Forward with the Pro EZGC Chromatogram Modeler
This improved version of our popular Pro EZGC chromatogram modeler for polymer capillary columns is just as simple to use as the original, but it now offers advanced options for selecting phases, changing carrier gases and control parameters, further optimizing your results, and much more! Already a favorite of analysts around the world, the updated software helps you develop new methods or optimize existing ones more efficiently and effectively than ever before.
In just seconds, you can generate a customized, interactive model chromatogram that matches real-world chromatograms with exceptional accuracy. Zoom in, view chemical structures, and even overlay the mass spectra of coeluting compounds.
Pro EZGC Chromatogram ModelerYOU NEED: To develop a method from scratch, including the column YOU HAVE: An analyte list (and you may have a column in mind, too). YOU GET: Customized, interactive model chromatograms that provide a specific phase, column dimensions and conditions. You can change columns, modify conditions, zoom in, view chemical structures, and even overlay mass spectra of coeluting compounds. Watch our instructional videos and get started today at www.restek.com/proezgc |
Modify Methods Quickly and with Confidence Using the EZGC Method Translator and Flow Calculator
The EZGC method translator and flow calculator tool makes it simple to switch carrier gases, change column dimensions or control parameters, or to optimize a method for speed or efficiency. Simply enter your method specifications, and the program will return a full set of calculated method conditions that will provide similar chromatography. Use the EZGC method translator and flow calculator tool to optimize your analysis for speed so you can increase sample throughput!
EZGC Method Translator and Flow CalculatorYOU NEED: To switch carrier gases, to change column dimensions or control parameters, or to optimize a method for speed or efficiency. YOU HAVE: An existing method. YOU GET: A full set of calculated method conditions that will provide similar chromatography. Results include oven program and run time as well as average velocity, flow rate, splitless valve time, and other control parameters—all in an easy-to-use, single-screen interface with seamless transfer between tools. Start saving time today—develop, optimize, or translate methods quickly and with confidence using Restek’s EZGC online software suite! www.restek.com/ezgc-mtfc |