PFAS Analysis by ASTM D8421-22 for Aqueous Samples

articleBanner
  • Sample Matrix: aqueous samples
  • Sample Preparation: filtration
  • Detection Technique: LC-MS/MS

Standard Test Method for Determination of Per- and Polyfluoroalkyl Substances (PFAS) in Aqueous Matrices by Co-solvation followed by Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS)

ASTM D8421-22 is used for the analysis of 44 PFAS in aqueous sample matrices. It combines filtration sample preparation with LC-MS/MS analysis. ASTM D8421-22 is not available from Restek, but it may be purchased from ASTM. The full analyte list and products recommended by Restek are provided below. In addition, you can register and directly compare analyte lists and matrices across PFAS analysis methods here or visit www.restek.com/PFAS to explore our technical resources for PFAS analysis.

 

Analyte List for ASTM D8421-22 for PFAS Analysis

PFAS acronyms, compound names, and CAS numbers vary across methods and guidelines, and they also depend on compound form (e.g., neutral vs. salt). The following list uses standardized terminology to allow easy comparison across the methods and guidelines in our comprehensive guide. Please refer to ASTM D8421-22 for method-specific nomenclature.

Acronym Compound Name CAS#
NEtFOSAA N-ethyl perfluorooctanesulfonamidoacetic acid 2991-50-6
NMeFOSAA N-methyl perfluorooctanesulfonamidoacetic acid 2355-31-9
PFBS Perfluoro-1-butanesulfonic acid 375-73-5
PFDA Perfluorodecanoic acid 335-76-2
PFDoA Perfluorododecanoic acid 307-55-1
PFHpA Perfluoroheptanoic acid 375-85-9
PFHxS Perfluoro-1-hexanesulfonic acid 355-46-4
PFHxA Perfluorohexanoic acid 307-24-4
PFNA Perfluorononanoic acid 375-95-1
PFOS Perfluorooctanesulfonic acid 1763-23-1
PFOA Perfluorooctanoic acid 335-67-1
PFTeDA Perfluorotetradecanoic acid 376-06-7
PFTrDA Perfluorotridecanoic acid 72629-94-8
PFUnA Perfluoroundecanoic acid 2058-94-8
HFPO-DA, GenX Hexafluoropropylene oxide dimer acid 13252-13-6
PFBA Perfluorobutanoic acid 375-22-4
PFPeA Perfluoropentanoic acid 2706-90-3
PFDS Perfluoro-1-decanesulfonic acid 335-77-3
PFHpS Perfluoro-1-heptanesulfonic acid 375-92-8
PFPeS Perfluoro-1-pentanesulfonic acid 2706-91-4
FOSA Perfluoro-1-octanesulfonamide 754-91-6
4:2 FTS 1H,1H,2H,2H-Perfluorohexane sulfonic acid 757124-72-4
6:2 FTS 1H,1H,2H,2H -Perfluorooctane sulfonic acid 27619-97-2
8:2 FTS 1H,1H,2H,2H -Perfluorodecane sulfonic acid 39108-34-4
8:2 FTUCA, FOUEA 2H-perfluoro-2-decenoic acid 70887-84-2
NMeFOSA N-Methylperfluorooctanesulfonamide 31506-32-8
NEtFOSA N-ethylperfluorooctanesulfonamide 4151-50-2
9Cl-PF3ONS 9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid (F-53B Major) 756426-58-1
11Cl-PF3OUdS 11-Chloroeicosafluoro-3-oxaundecane-1-sulfonic acid (F-53B Minor) 763051-92-9
ADONA 4,8-Dioxa-3H-perfluorononanoic acid 919005-14-4
PFNS Perfluoro-1-nonanesulfonic acid 68259-12-1
6:2 FHUEA 2H-perfluoro-2-octenoic acid 70887-88-6
PFEESA Perfluoro(2-ethoxyethane)sulfonic acid 113507-82-7
NFDHA Nonafluoro-3,6-dioxaheptanoic acid 151772-58-6
PFMPA Perfluoro-3-methoxypropanoic acid 377-73-1
PFMBA Perfluoro-4-methoxybutanoic acid 863090-89-5
NMeFOSE 2-(N-methylperfluoro-1-octanesulfonamido)-ethanol 24448-09-07
NEtFOSE 2-(N-ethylperfluoro-1-octanesulfonamido)-ethanol 1691-99-2
PFDoS Perfluorododecane sulfonate 79780-39-5
3:3 FTCA 3:3 Fluorotelomer carboxylic acid 356-02-5
5:3 FTCA 5:3 Fluorotelomer carboxylic acid 914637-49-3
7:3 FTCA, FHpPA 7:3 Fluorotelomer carboxylic acid or 3-perfluoroheptyl propanoic acid 812-70-4
LiTFSI (LiTFSA, HQ-115) Lithium Bis(trifluoromethane)sulfonimide 90076-65-6
PFPrA Pentafluoropropanoic acid 422-64-0

 

Recommended Products for ASTM D8421-22 for PFAS Analysis

ANALYSIS

Analytical LC Columns:

  • Raptor C18 SPP column: 2.7 μm, 50 x 2.1 mm (cat.# 9304A52)
  • Force C18 FPP column: 1.8 μm, 50 x 2.1 mm (cat.# 9634252)

Delay Columns:

  • PFAS delay column: 5 µm, 50 x 2.1 mm (cat.# 27854)
  • PFAS delay column: 5 µm, 50 x 3.0 mm (cat.# 580959 [custom product])

Guard Columns:

  • UltraShield UHPLC precolumn filter: 0.2 µm frit (cat.# 25809)
  • EXP direct connect holder for EXP guard cartridges (cat.# 25808)
  • Raptor C18 EXP guard column cartridge: 2.7 µm, 5 x 2.1 mm (cat.# 9304A0252)

SAMPLE HANDLING

Sample Prep:

  • 25 mm syringe filter: 0.22 µm, polypropylene (cat.# 28935)

GENERAL LAB SUPPLIES

  • Polypropylene vials: 100-pk. (cat.# 23243)
  • Polypropylene vials: 1000-pk. (cat.# 23246)
  • Polyethylene vial caps: 100-pk. (cat.# 23244)
  • Polyethylene vial caps: 1000-pk. (cat.# 23247)
  • Centrifuge tubes: 50 mL, 50-pk. (cat.# 25846)
  • Centrifuge tubes: 50 mL, 500-pk. (cat.# 28290)
  • Norm-Ject plastic syringe: 10 mL Luer Lock tip (cat.# 22775)*
  • Norm-Ject plastic syringe: 5 mL Luer Lock tip (cat.# 22774)*
    * Note that ASTM methods refer to the use of a glass syringe. Norm-Ject plastic syringes are made from polypropylene/polyethylene and can be used as an alternative.

 