PFAS Analysis by U.S. USDA CLG-PFAS 2.04 for Bovine, Porcine, Poultry, Siluriformes Muscle, and Bovine Plasma

articleBanner
  • Sample Matrix: bovine, porcine, poultry, and Siluriformes muscle and bovine plasma
  • Sample Preparation: protein precipitation extraction
  • Detection Technique: UHPLC-MS/MS

Screening, Determination, and Confirmation of PFAS by UHPLC-MS-MS

USDA Method CLG-PFAS 2.04 is used for the analysis of 16 PFAS in bovine, porcine, poultry, Siluriformes muscle samples, and in bovine plasma samples. It combines a methanol-based protein precipitation extraction with UHPLC-MS/MS analysis on a C18 column. The method is performance based and allows equivalent equipment, reagents, or solutions to be used, and some instrument parameters can be optimized as detailed in the method to improve chromatography and detection. The full analyte list and recommended products for PFAS analysis in bovine, porcine, poultry, Siluriformes muscle, and bovine plasma samples by CLG-PFAS 2.04 are provided below. In addition, you can register and directly compare analyte lists and matrices across PFAS analysis methods here or visit www.restek.com/PFAS to explore our technical resources for PFAS analysis.

 

Analyte List for USDA CLG-PFAS 2.04 for PFAS Analysis

PFAS acronyms, compound names, and CAS numbers vary across methods and guidelines, and they also depend on compound form (e.g., neutral vs. salt). The following list uses standardized terminology to allow easy comparison across the methods and guidelines in our comprehensive guide. Please refer to U.S. USDA CLG-PFAS 2.04 for method-specific nomenclature.

Acronym Compound Name CAS#
PFBS Perfluoro-1-butanesulfonic acid 375-73-5
PFDA Perfluorodecanoic acid 335-76-2
PFDoA Perfluorododecanoic acid 307-55-1
PFHpA Perfluoroheptanoic acid 375-85-9
PFHxS Perfluoro-1-hexanesulfonic acid 355-46-4
PFHxA Perfluorohexanoic acid 307-24-4
PFNA Perfluorononanoic acid 375-95-1
PFOS Perfluorooctanesulfonic acid 1763-23-1
PFOA Perfluorooctanoic acid 335-67-1
PFTeDA Perfluorotetradecanoic acid 376-06-7
PFTrDA Perfluorotridecanoic acid 72629-94-8
PFUnA Perfluoroundecanoic acid 2058-94-8
PFPeA Perfluoropentanoic acid 2706-90-3
PFDS Perfluoro-1-decanesulfonic acid 335-77-3
PFHxDA Perfluoro-n-hexadecanoic acid 67905-19-5
PFODA Perfluoro-n-octadecanoic acid 16517-11-6

 

Recommended Products for U.S. USDA CLG-PFAS for PFAS Analysis

ANALYSIS

Analytical LC Columns:

  • Raptor C18 SPP column: 2.7 μm, 50 x 2.1 mm (cat.# 9304A52)
  • Force C18 FPP column: 1.8 μm, 50 x 2.1 mm (cat.# 9634252)

Delay Columns:

  • PFAS delay column: 5 µm, 50 x 2.1 mm (cat.# 27854)
  • PFAS delay column: 5 µm, 50 x 3.0 mm (cat.# 580959 [custom product])

Guard Columns:

  • UltraShield UHPLC precolumn filter: 0.2 µm frit (cat.# 25809)
  • EXP direct connect holder for EXP guard cartridges (cat.# 25808)
  • Raptor C18 EXP guard column cartridge: 2.7 µm, 5 x 2.1 mm (cat.# 9304A0252)

SAMPLE HANDLING

Sample Prep:

  • Q-sep multispeed centrifuge for QuEChERS (cat.# 28295)

Sample Reservoirs:

  • Polypropylene sample reservoir: 75 mL (cat.# 26015)
  • Polypropylene connectors (cat.# 26007)

Vacuum Manifolds:

  • Resprep QR-24 vacuum manifold (cat.# 28299-VM)
  • Resprep QR-12 vacuum manifold (cat.# 28298-VM)

Vacuum Pumps:

  • Chemker 300 PTFE vacuum pump: 20 L/min, AC110 V, 60 Hz (cat.# 27426)
  • Chemker 300 PTFE vacuum pump: 18 L/min, AC220 V, 50 Hz (cat.# 27427)

GENERAL LAB SUPPLIES

  • Polypropylene vials: 100-pk. (cat.# 23243)
  • Polypropylene vials: 1000-pk. (cat.# 23246)
  • Polyethylene vial caps: 100-pk. (cat.# 23244)
  • Polyethylene vial caps: 1000-pk. (cat.# 23247)
  • Centrifuge tubes: 50 mL, 50-pk. (cat.# 25846)
  • Centrifuge tubes: 50 mL, 500-pk. (cat.# 28290)
  • Norm-Ject plastic syringe: 10 mL Luer Lock tip (cat.# 22775)
  • Norm-Ject plastic syringe: 5 mL Luer Lock tip (cat.# 22774)

 

Discover More Solutions for PFAS Analysis!

 