Can HPLC-UV Be Used For Terpenes Analysis In Cannabis?

28 Dec 2016

While HPLC may be tempting to use for terpenes analysis, a GC/FID or GC/MS is really the most straightforward and recommended way of analyzing terpenes in cannabis. Terpenes, being relatively volatile and neutral, lend themselves rather nicely to GC in general.

As you can see from Figure 1 below, coelutions of the cannabinoids and terpenes are very likely when analyzing real cannabis samples by HPLC-UV methods. A Shimadzu Prominence 20AD HPLC system and Raptor ARC-18, 2.7um, 150 x 4.6mmID column (catalog# 9314A65) was used in Figure 1.

Figure 1- Standards at equal concentrations
figure-1-blog-terpenes

At 205nm, there are many co-elutions of terpenes and the cannabinoids, making identification and quantitation extremely difficult, if not impossible, for either class of compounds.

Conversely, cannabinoids can be analyzed by HPLC-UV, as long as the correct UV wavelength is chosen. At 220nm, terpenes yield a very low signal. Because cannabinoids of interest are present at a much higher concentration than terpenes, in addition to providing a stronger UV signal at 220nm, cannabinoids can be reliably analyzed at a wavelength of 220nm. See Figure 2.

Figure 2-

Column Raptor ARC-18 (cat.# 9314A65)
Dimensions: 150 mm x 4.6 mm ID
Particle Size: 2.7 µm
Pore Size: 90 Å
 
Guard Column: Raptor ARC-18 EXP guard column cartridge, 2.7 µm (cat.# 9314A0250)
Temp.: 30 °C
Standard/Sample Tetrahydrocannabivarin (cat.# 34100)
  Cannabidiolic acid (CBDA) (cat.# 34094)
  Cannabichromene (CBC) (cat.# 34092)
  Cannabigerol (CBG) (cat.# 34091)
  delta-9-Tetrahydrocannabinolic acid A (THCA-A) (cat.# 34093)
  delta-8-Tetrahydrocannabinol (Δ8-THC) (cat.# 34090)
  delta-9-Tetrahydrocannabinol (Δ9-THC) (cat.# 34067)
  Cannabinol (CBN) (cat.# 34010)
  Cannabidiol (CBD) (cat.# 34011)
  Compounds not present in these mixes were obtained separately.
Diluent: 25:75 Water:methanol
Conc.: 50 µg/mL
Inj. Vol.: 5 µL
Mobile Phase  
A: Water, 5 mM ammonium formate, 0.1% formic acid
B: Acetonitrile, 0.1% formic acid
 
Time (min) Flow (mL/min) %A %B
0.00 1.5 25 75
9.00 1.5 25 75
Detector UV/Vis @ 228 nm
Instrument HPLC

 

In summary, HPLC-UV analysis of terpenes in cannabis is not recommended, and will likely cause more issues than it will provide solutions.

A good solution to the coelutions by HPLC-UV is to choose a GC headspace method. Interferences from the complex sample matrix, as well as the much less volatile cannabinoids can be eliminated then.

 