Ethylene Oxide in Air Canisters – Method Performance
10 May 2021
In previous posts on ethylene oxide (EtO) I've talked about the following:
- Resolving interferences with cryogenic cooling
- How to resolve EtO and the TO-15A suite of compounds
- Canister storage issues with EtO
In this post we’ll cap it all off with the method performance data I was able to obtain, including injection volume, linearity, calibration curves, detection limits, and precision.
First, let’s have a brief refresher on the method parameters used. Table 1 shows the GC-MS and preconcentrator settings used for the method.
Preconcentrator | Markes Unity 2 + Kori + CIA |
Unity 2 Settings | |
Unity Trap Low | −30 °C |
Desorb temp | 300 °C |
Desorb flow | 6 mL/min |
Desorb time | 180 s |
Desorb Split Flow | 3 mL/min |
Flow Path Temperature | 120 °C |
Internal Standard | |
Purge flow | 50 mL/min |
Purge time | 60 s |
Volume | 50 mL |
ISTD flow | 50 mL/min |
CIA Advantage Settings | |
Volume | 400 mL |
Purge flow | 50 mL/min |
Purge time | 60 s |
Sample flow | 100 mL/min |
Kori-xr Settings | |
Kori Trap Low | −5 °C |
Kori Trap High | 300 °C |
GC | Agilent 7890B |
Injection type | On-column |
Column | 624Sil MS 60 m × 0.25 mm × 1.4 µm |
Carrier gas | He, constant flow |
Flow rate | 2 mL/min |
Oven temp |
0 °C (hold 5 min) to 60 °C at 3.5 °C/min |
Detector | MS Agilent 5977A |
Acquisition mode | SIM/Scan |
Scan parameters | |
Scan range (amu) | 29-226 |
Scan rate (scans/s) | 3.7 |
SIM parameters | |
SIM ions | 15, 29, 43, 44, 56 |
Dwell time | 50 |
Transfer line | 250 °C |
Analyzer type | Quadrupole |
Source type | Extractor |
Source temp | 350 °C |
Quad temp | 200 °C |
Electron energy | 70 eV |
Solvent delay time | 0 min |
Tune type | BFB |
Ionization mode | EI |
Table 1: Instrument settings for combined EtO and TO-15A analysis. The sample is split at the preconcentrator during desorb.
All standards and blanks used in these studies were made in 50% relative humidity (%RH) zero air using 6 L SilcoCans. My previous post on canister storage for EtO shows the importance of using humid air when testing for EtO, showing that dry air can potentially hide EtO interferences in real world humid samples.
Once the initial settings were determined, we had to ensure that we wouldn’t be overloading the trap at higher concentrations. To do this we took two concentrations of standard and increased the volume loaded onto the preconcentrator to see if there was any fall off at larger volumes. We found that up to 600 mL at 2,688 pptv the system was linear, as shown in Figure 1. As a best balance between sensitivity and sample load time, 400 mL was used as the standard injection volume for our study.
Figure 1: EtO linearity with preconcentrator volume
Once it was established that we were in the linear range for EtO, a calibration from 34 pptv to 2,688 pptv was made using bromochloromethane as an internal standard. TO-15A states that a calibration should have less than 30% RSD between the relative response factors of the calibration points, and each point must be within 20% of the true value. As shown in Figure 2, the %RSD was 12.8%, and Table 2 shows the calibration points are all within ±20% of their true value.
Figure 2: EtO calibration results
True (pptv) |
34 |
67 |
134 |
269 |
672 |
1344 |
2688 |
Calculated (pptv) |
40 |
58 |
150 |
255 |
727 |
1196 |
2446 |
% from true |
119% |
86% |
112% |
95% |
108% |
89% |
91% |
Table 2: Calculated recovery of EtO calibration standards
With the calibration established it was time to determine the method detection limit (MDL) of the method. Seven replicates at the low calibration point were analyzed, and the standard deviation was multiplied by the student T value of 3.143 to calculate the MDL. The limit of quantitation (LOQ) was taken as 3 times the MDL. As shown in Table 3 we were able to get relatively low into the pptv range for our detection limits.
Replicate |
1 |
2 |
3 |
4 |
5 |
6 |
7 |
Standard Deviation |
MDL (pptv) |
LOQ (pptv) |
EtO (pptv) |
41 |
43 |
38 |
34 |
45 |
33 |
50 |
5.6 |
18 |
55 |
Table 3: EtO MDL and LOQ results
Finally, the precision and accuracy of the method was determined by analyzing 4 replicates at 500 pptv. In addition, a stability study was done by analyzing 4 replicates over the course of over 2 weeks to determine how long EtO can be kept in canisters. Tables 4 and 5 show these results.
Replicate |
1 |
2 |
3 |
4 |
Average |
SD |
RSD |
EtO (pptv) |
514 |
417 |
588 |
560 |
520 |
65 |
13% |
% recovery |
103% |
83% |
118% |
112% |
104% |
Table 4: EtO precision and accuracy
Replicate |
1 |
2 |
3 |
4 |
Average |
Day 1 (% recovery) |
100% |
122% |
82% |
88% |
98% |
Day 2 (% recovery) |
109% |
129% |
101% |
100% |
110% |
Day 5 (% recovery) |
97% |
87% |
106% |
100% |
108% |
Day 8 (% recovery) |
96% |
119% |
130% |
110% |
114% |
Day 12 (% recovery) |
137% * |
138%* |
134% * |
122% |
133% * |
Day 16 (% recovery) |
107% |
132%* |
132% * |
116% |
126% |
Table 5: EtO stability. Results flagged with * fail the TO-15A ±30% recovery
To sum everything up, by using cryogenic cooling EtO can be separated from interferences and combined with TO-15A with detection limits down to 18 pptv, with good accuracy and precision, and with stability of a least 1 week in SilcoCans. Given what we’ve found with EtO growth in canisters any lab testing for EtO should test their own canisters to determine a realistic holding time for their samples.
For a more in depth look at EtO analysis, check out the article by myself and Jason Herrington that was recently published in Separations.
https://www.mdpi.com/2297-8739/8/3/35
I also gave a presentation at NEMC 2020 on EtO stability in air canisters.