Potency Testing by LC-MS, the Future of Cannabinoid Testing?
29 Apr 2021Potency testing is the bread and butter of the cannabis industry. This analysis is typically performed by LC-UV as this instrumentation is relatively inexpensive and requires minimal training. There are over 150 cannabinoids that have been discovered thus far, with more likely to be discovered in the coming years as cannabis research continues to accelerate. While LC-UV is a robust and reliable approach, it has some limitations. LC-UV is hindered by the need to resolve all analytes and lacks sensitivity when compared to MS. As new cannabinoids are discovered and gain interest, additional cannabinoids will need to be added into potency testing methods. MS can offer a sensitive and selective analytical solution which enables lower limits of detection and eliminates the need to resolve compounds that are not isobars. Here, a method has been developed for the analysis of 21 cannabinoids by LC-MS in 9 minutes that also achieves the resolution of all isobaric compounds.
Column |
Raptor ARC-18 (cat.#9314A62) |
|
Dimensions: |
150 mm x 2.1 mm ID |
|
Particle Size: |
2.7 µm |
|
Guard Column: |
Raptor ARC-18 EXP guard column cartridge 2.7 µm (cat.#9314A0252) |
|
Temp.: |
30 °C |
|
Conc.: |
500 ppb |
|
Inj. Vol.: |
2 µL |
|
Mobile Phase |
|
|
A: |
Water, 0.1% formic acid |
|
B: |
Acetonitrile & Methanol 50/50, 0.1% formic acid |
|
Time (min) |
Flow (mL/min) |
%B |
0 |
0.5 |
80 |
0.9 |
0.5 |
80 |
Ion Mode |
ESI (+)/(-) |
|
Detector |
MS |
Peaks |
SIM |
Ion Mode |
Cannabidivarin |
287 |
+ |
Cannabidivarinic Acid |
331 |
+ |
Cannabidiol |
315 |
+ |
Cannabigerol |
317 |
+ |
Cannabidiolic Acid |
357 |
- |
Tetrahydrocannabivarin |
287 |
+ |
Cannabigerolic Acid |
359 |
- |
Cannabichromevarin |
287 |
+ |
Cannabinol |
311 |
+ |
Cannabidiphorol |
343.5 |
+ |
Tetrahydrocannabivarinic Acid |
331 |
+ |
Δ9-Tetrahydrocannabinol |
315 |
+ |
Δ8-Tetrahydrocannabinol |
315 |
+ |
Cannabinolic Acid |
353 |
- |
Cannabicyclol |
315 |
+ |
Cannabichromene |
315 |
+ |
Tetrahydrocannabinolic Acid-A |
357 |
- |
Cannabicitran |
315 |
+ |
Tetrahydrocannabiphorol |
343.5 |
+ |
Cannabichromenic Acid |
357 |
- |
Cannabicyclolic acid |
357 |
- |
In the above compound table, each analyte is listed with their SIM and ESI ion mode. All isobars are color coded and these compounds will need to be chromatographically separated.
In this method, 21 cannabinoids are capable of being analyzed by LC-MS. By using MS detection, some of the burden is lifted to chromatographically separate all compounds, as only isobars need to be separated. Some of the isobars are more difficult to separate than others. The most challenging set to separate are the m/z of 315, which consists of THC and its isobars.
By using this method, the desired resolution can be achieved for key compounds, including THC isobars. Additional cannabinoids can be added to the analyte list as more are discovered with minimal further optimization. Could MS be the future of cannabinoid potency testing?