27 Nov 2019

We recently took a look at combining Vitamin E Acetate and cannabinoids into one method.  This route allows labs to diversify their testing services with a few simple modifications, if using our potency method.  We wanted to give labs another option as well.  Therefore, Justin Steimling went back into the lab and developed a standalone rapid analysis for Vitamin E Acetate.


graphical user interface

A Raptor ARC-18 30 x 3.0 mm, 2.7 µm column was utilized for this quick 3 minute run.  Justin used methanol instead of acetonitrile since it is actually a stronger solvent for this compound.  Another added benefit to this method is the lower cost associated with methanol compared to acetonitrile.  Whether your lab would like a rapid standalone method or modifying your potency method, we have two routes for analyzing Vitamin E Acetate.