You want them, I got them...a lot of them: pesticides GC-MS/MS transitions

20 May 2015

gcmsmsJAK15
I can tell you that determining transitions (SRMs) for GC-MS/MS can be quite challenging with so many precursor ions to evaluate. I spent a considerable amount of time optimizing transitions for our GC Multiresidue Pesticide Kit which has 203 compounds. That is a lot of transitions to evaluate!

I also tested my chromatographic method for analyte to analyte interferences by comparing ion ratios. The ion ratio of qualifying Q3 ion to quantification Q3 ion for each compound tested in small subsets was compared to the ion ratio determined in the presence of all 203 compounds. When the two ratios fell outside of plus/minus thirty percent, this indicated analyte to analyte interference and alternative transitions were used.

Transitions for all compounds and notes about analyte to analyte interferences and shared transitions can be found in this downloadable spreadsheet. The data was collected and optimized on our Thermo Scientific TSQ 8000 Triple Quadrupole GC-MS System.

The spreadsheet provides the method conditions (last tab), retentions times and three SRMs per compound. CAS numbers and isomers information is also included.

I hope that this helps accelerate your method development!