Sulfur Compounds on Rxi-1ms
Peaks | |
---|---|
1. | Hydrogen sulfide![]() |
2. | Carbonyl sulfide![]() |
3. | Methyl mercaptan![]() |
4. | Ethyl mercaptan![]() |
5. | Dimethyl sulfide![]() |

GC_AR00861
Column | Rxi-1ms, 30 m, 0.32 mm ID, 4.00 µm (cat.# 13396) |
Sample | Hydrogen sulfide, carbonyl sulfide, methyl mercaptan, ethyl mercaptan, dimethyl sulfide |
Diluent: | Helium |
Conc.: | 100 ppbv |
Injection | splitless |
Inj. Temp.: | 30 °C |
Oven | |
Oven Temp.: | 30 °C |
Carrier Gas | He, constant pressure |
Linear Velocity: | 48 cm/sec @ 30 °C |
Detector | SCD @ 800 °C |
Notes | Injection: sample loop (30°C), 1 mL splitless direct Sample Storage and Tansfer: SilcoCan air monitoring canister with Siltek-treated 1/4" valve (cat.# 24182-650); Sulfinert-treated gas sample loop, 1 cc (cat.# 22848); Sulfinert-treated gas sample loop, 10 cc (custom order) |