delim
delim
TO-15 65 Component Mix on Rtx-VMS (30 m, 1.2 mL/min)
This chromatogram may contain analytes that are synonyms of your search terms.

TO-15 65 Component Mix on Rtx-VMS (30 m, 1.2 mL/min)

PeakstR (min)
1.Propylene1.98
2.Dichlorodifluoromethane (Freon® 12)2.03
3.1,2-Dichlorotetrafluoroethane (Freon® 114)2.19
4.Chloromethane2.24
5.Vinyl chloride2.34
6.1,3-Butadiene2.36
7.Acetaldehyde (contaminant)2.64
8.Bromomethane2.68
9.Chloroethane2.81
10.Trichlorofluoromethane (Freon® 11)2.96
11.1,1-Dichloroethene3.45
12.Ethanol3.46
13.Carbon disulfide3.46
14.1,1,2-Trichlorotrifluoroethane (Freon® 113)3.50
15.Acrolein3.79
16.Isopropyl alcohol3.99
17.Methylene chloride4.01
18.Acetone4.07
19.trans-1,2-Dichloroethene4.16
20.Acetonitrile (contaminant)4.24
21.Hexane4.24
22.Methyl tert-butyl ether (MTBE)4.28
23.1,1-Dichloroethane4.76
24.Vinyl acetate5.03
25.cis-1,2-Dichloroethene5.30
26.Cyclohexane5.48
27.Bromochloromethane (IS)5.49
28.Chloroform5.58
29.Carbon tetrachloride5.70
30.Ethyl acetate5.72
31.Tetrahydrofuran5.74
32.1,1,1-Trichloroethane5.78
33.2-Butanone (MEK)5.91
34.Heptane6.13
35.Benzene6.18
PeakstR (min)
36.1,2-Dichloroethane6.41
37.Trichloroethylene6.82
38.1,4-Difluorobenzene (IS)6.86
39.1,2-Dichloropropane7.43
40.Bromodichloromethane7.50
41.Methyl methacrylate7.71
42.1,4-Dioxane7.76
43.cis-1,3-Dichloropropene8.28
44.Toluene8.58
45.Tetrachloroethene9.07
46.4-Methyl-2-pentanone (MIBK)9.12
47.trans-1,3-Dichloropropene9.15
48.1,1,2-Trichloroethane9.36
49.Dibromochloromethane9.59
50.1,2-Dibromoethane9.91
51.2-Hexanone (MBK)10.29
52.Chlorobenzene-d5 (IS)10.65
53.Chlorobenzene10.67
54.Ethylbenzene10.73
55.m-Xylene10.96
56.p-Xylene10.96
57.o-Xylene11.61
58.Styrene11.70
59.Bromoform11.71
60.4-Bromofluorobenzene*12.54
61.1,1,2,2-Tetrachloroethane12.92
62.4-Ethyltoluene12.98
63.1,3,5-Trimethylbenzene13.13
64.1,2,4-Trimethylbenzene13.77
65.1,3-Dichlorobenzene14.27
66.1,4-Dichlorobenzene14.43
67.Benzyl chloride14.90
68.1,2-Dichlorobenzene15.15
69.Hexachlorobutadiene17.28
70.1,2,4-Trichlorobenzene17.29
71.Naphthalene17.62
*Tuning standard
image
GC_AR1155
ColumnRtx-VMS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 19915)
SampleTO-15 65 component mix (cat.# 34436)
TO-14A internal standard/tuning mix (cat.# 34408)
Diluent:Nitrogen
Conc.:10.0 ppbv 400 mL injection
InjectionDirect
Oven
Oven Temp.:32 °C (hold 1 min) to 150 °C at 8 °C/min to 230 °C at 33 °C/min
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
Linear Velocity:40 cm/sec @ 32 °C
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1035-2503.32
Transfer Line Temp.:230 °C
Analyzer Type:Quadrupole
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:69.9 eV
Solvent Delay Time:1.0 min
Tune Type:BFB
Ionization Mode:EI
PreconcentratorNutech 8900 DS
  Trap 1 Settings
Type/Sorbent :Glass beads
Cooling temp:-155 °C
Preheat temp:5 °C
Preheat time:0 sec
Desorb temp:20 °C
Desorb flow:5 mL/min
Desorb time:360 sec
Bakeout temp:200 °C
Flush flow:120 mL/min
Flush time:60 sec
Sweep flow:120 mL/min
Sweep time:60 sec
  Trap 2 Settings
Type/Sorbent:Tenax
Cooling temp:-35 °C
Desorb temp:190 °C
Desorb time:30 sec
Bakeout temp:200 °C
Bakeout time:10 sec
  Cryofocuser
Cooling temp:-160 °C
Inject time:140 sec
  Internal Standard
Purge flow:100 mL/min
Purge time:6 sec
Vol.:100 mL
ISTD flow:100 mL/min
  Standard
Size:200 mL
Purge flow:100 mL/min
Purge time:6 sec
Sample flow:100 mL/min
InstrumentHP6890 GC & 5973 MSD
AcknowledgementNutech
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

TO-15

,

Air

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us