share:
delim
delim
NJ Low Level TO-15 75 Component Mix on Rtx-VMS (60 m, 40 °C start)

NJ Low Level TO-15 75 Component Mix on Rtx-VMS (60 m, 40 °C start)

PeakstR (min)
1.Propylene3.45
2.Dichlorodifluoromethane (Freon 12)3.54
3.1,2-Dichlorotetrafluoroethane (Freon 114)3.82
4.Chloromethane3.94
5.n-Butane4.00
6.Vinyl chloride4.11
7.1,3-Butadiene4.15
8.Bromomethane4.81
9.Chloroethane5.09
10.Vinyl bromide5.35
11.n-Pentane5.39
12.Trichlorofluoromethane (Freon 11)5.40
13.1,1-Dichloroethene6.58
14.Ethanol6.61
15.Carbon disulfide6.63
16.1,1,2-Trichlorotrifluoroethane (Freon 113)6.69
17.Acrolein7.45
18.Allyl chloride7.68
19.Isopropyl alcohol7.87
20.Methylene chloride7.87
21.Acetone8.03
22.trans-1,2-Dichloroethene8.13
23.Hexane8.25
24.Methyl tert-butyl ether (MTBE)8.37
25.Tertiary butanol8.58
26.1,1-Dichloroethane9.01
27.Vinyl acetate9.32
28.cis-1,2-Dichloroethene9.63
29.Cyclohexane9.82
30.Bromochloromethane (IS)9.84
31.Chloroform9.90
32.Carbon tetrachloride10.03
33.Ethyl acetate10.04
34.1,1,1-Trichloroethane10.10
35.Tetrahydrofuran10.10
36.2-Butanone (MEK)10.23
37.Isooctane10.29
38.Heptane10.37
39.Benzene10.44
PeakstR (min)
40.1,2-Dichloroethane10.62
41.Trichloroethylene10.91
42.1,4-Difluorobenzene (IS)10.93
43.1,2-Dichloropropane11.32
44.Bromodichloromethane11.35
45.Methyl methacrylate11.44
46.1,4-Dioxane11.51
47.cis-1,3-Dichloropropene11.78
48.Toluene11.93
49.4-Methyl-2-pentanone (MIBK)12.17
50.Tetrachloroethene12.18
51.trans-1,3-Dichloropropene12.18
52.1,1,2-Trichloroethane12.30
53.Dibromochloromethane12.42
54.1,2-Dibromoethane12.59
55.2-Hexanone (MBK)12.67
56.n-Nonane12.74
57.Ethylbenzene12.87
58.Chlorobenzene-d5 (IS)12.88
59.Chlorobenzene12.89
60.m-Xylene12.95
61.p-Xylene12.95
62.o-Xylene13.21
63.Styrene13.24
64.Bromoform13.30
65.Cumene13.38
66.4-Bromofluorobenzene*13.57
67.n-Propyl benzene13.61
68.1,1,2,2-Tetrachloroethane13.65
69.4-Ethyltoluene13.66
70.1,3,5-Trimethylbenzene13.69
71.2-Chlorotoluene13.73
72.1,2,4-Trimethylbenzene13.93
73.1,3-Dichlorobenzene14.16
74.1,4-Dichlorobenzene14.21
75.Benzyl chloride14.34
76.1,2-Dichlorobenzene14.46
77.Hexachlorobutadiene15.29
78.1,2,4-Trichlorobenzene15.35
79.Naphthalene15.60
*Tuning standard
image
GC_AR1167
ColumnRtx-VMS, 60 m, 0.25 mm ID, 1.40 µm (cat.# 19916)
with MXT low-dead-volume connector kit (cat.# 20536)
SampleTO-14A internal standard/tuning mix (cat.# 34408)
75 comp TO15 + NJ mix (cat.# 34396)
Diluent:Air @ 50% RH
Conc.:5 ppbv 250 mL injection
Injectionon-column
Oven
Oven Temp.:40 °C (hold 7 min) to 250 °C at 30 °C/min (hold 2 min)
Carrier GasHe, constant flow
Flow Rate:2.0 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
10.0035.0 - 226.003.8
Transfer Line Temp.:250 °C
Analyzer Type:Quadrupole
Source Type:Extractor
Extractor Lens:6 mm ID
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70.0 eV
Tune Type:BFB
Ionization Mode:EI
PreconcentratorMarkes CIA Advantage
InstrumentAgilent 7890B GC & 5977A MSD
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

TO-15

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us