share:
delim
delim
Ethanol and IPA on Rtx-VMS (60 m, 40 °C start)
This chromatogram may contain analytes that are synonyms of your search terms.

Ethanol and IPA on Rtx-VMS (60 m, 40 °C start)

PeakstR (min)
1.Ethanol6.61
2.Isopropyl alcohol7.87
image
GC_AR1168
ColumnRtx-VMS, 60 m, 0.25 mm ID, 1.40 µm (cat.# 19916)
with MXT low-dead-volume connector kit (cat.# 20536)
SampleTO-14A internal standard/tuning mix (cat.# 34408)
75 comp TO15 + NJ mix (cat.# 34396)
Diluent:Air @ 50% RH
Conc.:5 ppbv 250 mL injection
Injectionon-column
Oven
Oven Temp.:40 °C (hold 7 min) to 250 °C at 30 °C/min (hold 2 min)
Carrier GasHe, constant flow
Flow Rate:2.0 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
10.0035.0 - 226.003.8
Transfer Line Temp.:250 °C
Analyzer Type:Quadrupole
Source Type:Extractor
Extractor Lens:6 mm ID
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70.0 eV
Tune Type:BFB
Ionization Mode:EI
PreconcentratorMarkes CIA Advantage
InstrumentAgilent 7890B GC & 5977A MSD
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

TO-15

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us